Button
Release of "Adjustment" feature for changes and calibrations. The header "Adjustment
active
" will appear.
Saving of entered values for
sensor
characteristics
.
Saving of entered values – and calibrations where applicable – from the temporary memory
buffer on the device.
Exiting this menu without taking over the entered values; the current data will be retained.
An adjusted characteristic is deleted and the factory default setting adopted. The appliance
needs the be re-adjusted!
Table 5
– Buttons Adjustment
Sensor characteristics
Indication of characteristics for sensors a to d at maximum anticipated load signal.
Load calibration
Calibration of sensors a to d takes place here.
Note:
No functioning of the LS 804 will be possible unless the sensors (load calibration) are
adjusted. Incorrect adjustment or no adjustment at all will unleash the following message:
Adj. not
OK!
"
Further information on adjustment status will be found under 6.3.1 Adjustment status !
Caution!
Adjusted values (load calibration) will not be taken over by the LS 804 until the SAVE
button is pressed! As long as the values have not been taken over the following message will
appear: "
Adj. not saved!"
!
Adjustment status
Refer to 6.3.1 Adjustment status.