© Associated Research 2017
59
MEASUREMENT
Range
Resolution
Accuracy
AC Current, A
0.00-40.00
0.01
± (3% of r 1 counts)
DC Current, A
0.00-40.00
0.01
± (3% of r 1 counts)
Resistance, mΩ
0-600
1
± (3% of r 3 counts) for 1 - 5.99A,
± (2% of r 2 counts) for 6 - 40A
AC Voltage, V
0.00-8.00
0.01
± (2% of r 2 counts)
DC Voltage, V
0.00-8.00
0.01
± (2% of r 2 counts)
GENERAL SPECIFICATIONS
Safety Agency Listing
CE, TUV SUD, RoHS2
PLC Remote Control
Input: Test, Reset, Interlock, Withstand Processing, Recall File 1 through 3(Standard), Hardware
Interlock- Direct control sine wave and amplifier relay off when Interlock signal is disable.
Output:- Pass, Fail, Test-in-Process, Reset-out, Start-out
(Output relay contact ratings is 1AAC/125VAC
Memory
50 steps
Internal memory- 1500 test results
Display
4.3” Color Display (Touch Panel)
Multinational language
User interface available in English, Traditional Chinese, Simplified Chinese, Turkish, Portuguese,
Spanish, German, French
Interface
Standard: USB
Security (Standard Password
Protection)
Programmable User ID and Password capability to avoid unauthorized access to test set-up
program.
Alarm Volume Setting
Range: 0 - 9 ; 0 = OFF, 1 is softest volume, 9 is loudest volume
Calibration
Software and adjustments are made through the front panel. Automatic Calibration alert
function to signal operator when the next calibration is due.
Environment
Temperature: 0 - 40°C, Humidity: 20 - 80%
Dimension
215mm(W) ×88.1mm(H) × 300 mm(D)
Weight
5kg