PREVENTATIVE MAINTENANCE:
The following preventative maintenance procedures should be carried out every
12 months or as otherwise specified:
• Overall functional check that all hand control functions are operating the bed as
expected.
• Overall inspection for visible signs of wear, tear or damage to the integrity of
electronics and Cables.
• Actuators and all moving components should not make unusual noises.
To prolong the working life of the bed and for trouble free operation it is recommended
that routine inspections should be performed by a person who is suitably qualified and
well acquainted with the design, use and care of ComfiMotion beds.
If inspection reveals that the safety of the bed is jeopardised in any way, it should be
IMMEDIATELY WITHDRAWN FROM SERVICE
, tagged as such and the owner notified,
until repair or replacement has occurred.
INSPECTION:
• A visual and functional inspection should be performed monthly
• Visually inspect electrical components: can you see damage to the hand control
or power cord?
• Visually inspect mechanical components: can you see damage, chips
or deformation to the frame, headboard or accessories?
• Is the hand control functioning as it should?
• Are the castors, brakes and accessories functioning as they should?
NOTE:
If you are unsure if your ComfiMotion bed is working correctly, Contact
Aidacare or your local authorised Aspire Dealer.
NOTE:
Other than general maintenance activities all other service or repair work must
be performed by qualified and trained persons or agents approved by Aidacare.
11. SERVICE & MAINTENANCE
Effect
Possible Cause
Verification
Corrective Action
BED DOES
NOT STAY IN
PLACE
Locking castors
not preventing
bed movement.
Brake System is
malfunctioning.
Not all locking castors
have been engaged.
Brake pedal is in lowest
position, but bed does
not stay in place.
Engage 2 x locking castors on the
foot end of the bed (both sides)
If this does not work, contact your
sales representative or a qualified
Aidacare service technician.
BED DOES
NOT MOVE
UP OR DOWN
Bed is not
plugged in and
optional battery
is flat.
Actuator is faulty.
Handset is faulty.
Check if the power lead
is plugged into mains
power.
Swap actuators.
Swap handsets.
If unplugged, plug power leads
into mains power. Allow 3 hours
to charge battery (if fitted).
Contact your sales representative
or a qualified Aidacare service
technician.
BACKREST
DOES NOT
MOVE UP OR
DOWN
Actuator is
unplugged.
Actuator is faulty.
Handset is faulty.
Check if cables have
come loose from control
box (under top frame at
head end).
Swap Actuator.
Swap handsets.
If loose, fully plug cables into
control box.
Contact your sales representative
or a qualified Aidacare service
technician.
BED DOES
NOT MOVE
AS EXPECTED
WHEN
PRESSING
HANDSET
BUTTONS
Control box
malfunction.
Actuators responding
to button presses on
handset but movements
do not seem correct.
Conduct an Electronics System
Rest. Conduct Factory Reset
if error still present.