background image

– 14 –

Step 1:

Press the CONFIG key until the word ZERO
SP appears.

Step 2:

Press ENTER.

Step 3:

Enter user five digit password (PW). This is
the same password established to access
the CONFIG mode in the menu.

Step 4:

Press the CONFIG key to select the 
desired option.

Step 5:

Press ENTER to finalize the selection.

Notes: 

Selecting the DISAB feature does not disable the
CLEAR button on the keypad for the MAX/MIN fea-

ture. If the DISAB feature is selected, pressing the
ZERO button on the keypad will cause the display to
read DISAB for two seconds. The gauge will then
revert back to the unit of measure of the gauge. The
DISAB feature disables the zero feature of the gauge. 

Zero Disable Feature

:

This feature allows disabling the Zero/Clear button
on the keypad. It also allows for a zero tolerance of
either 5% (default) or 10% of the gauge range.

Step 1:

Press the CONFIG key until the word ZERO
SP appears.

Step 2:

Press ENTER.

Step 3: 

A prompt appears to enter PW  (enter
password). The ZERO SP password is the
same password as discussed on page 10
and the heading CALIBRAT:, Gauge
Calibration.  Follow the instructions on
page 10 to enter a password.

Step 4:

Press the CONFIG key to select the zero
tolerance, either 5% or 10% of range, or

press the CONFIG key again and the word
DISAB appears. Press ENTER to select the
new default setting.

KEYPAD FUNCTIONS

I&M008 10080 Rev. E_layout  10/14/13  9:35 AM  Page 14

Содержание Digital Test Gauge

Страница 1: ...Ashcroft Inc 2013 I M008 10080 10 13 Rev E GBE Digital Test Gauge Operating Instructions...

Страница 2: ...2...

Страница 3: ...stry of 66 high Other industry leading fea tures include twelve selectable engineering units seven languages and password protected disable and calibration functions With the range printed on the keyp...

Страница 4: ...4...

Страница 5: ...ions CONFIG Mode Units Engineering 9 Update Rate 9 10 Auto Off 10 Backlite 10 Languages 10 Damping 11 Contrast 11 Calibrate Gauge Calibration 11 Default Password 12 Zero Calibration 13 Zero Span 13 Ze...

Страница 6: ...r maximum displayed values UNITS Pressure ENGLISH PSI INHG INH2O with Temperature Options 60 F 4 C 20 C FTSW METRIC BAR MBAR KPA MPA mmHG CMH2O mmH2O KG CM2 UPDATE RATE Pressure measurement per second...

Страница 7: ...charge Flashing display when unit pressured beyond full scale Press to turn backlite on or off Range on keypad complies with ASME B40 7 Press to turn unit on or off AUTO OFF Turns unit off after optio...

Страница 8: ...auge To clear minimum and maximum values press ZERO CLR button when min max values are indi cated This key allows for customization of the gauge Pressing the CONFIG key allows cycling through the main...

Страница 9: ...g the rate at which pressure is updated on the display screen The default rate measures pressure at the maximum rate of 10 updates per second or 100 milli seconds Optional rates of measurement are mea...

Страница 10: ...g back lite button on will decrease battery life To use the BACKLITE option Step 1 Press the CONFIG key until the word BACKLITE appears Step 2 Press the ENTER key Step 3 Press the CONFIG key to select...

Страница 11: ...1 2 3 4 5 6 and 7 Step 1 Press the CONFIG key until the word CONTRAST appears Step 2 Press the ENTER key Step 3 Press the CONFIG key to select one of the available CONTRAST options Step 4 Press the E...

Страница 12: ...rd Step 10 Press the ENTER key Step 11 Decide on a five number user password then follow the procedure above insert ing a number in the flashing display until all five numbers are inserted Step 12 A S...

Страница 13: ...djustment of full scale range an ERROR message will be displayed when the ENTER button is pressed If this occurs press the ZERO CLEAR button on the keypad to return to the previous screen 2 ASME B40 7...

Страница 14: ...back to the unit of measure of the gauge The DISAB feature disables the zero feature of the gauge Zero Disable Feature This feature allows disabling the Zero Clear button on the keypad It also allows...

Страница 15: ...tep 2 Press the ENTER key Step 3 Insert the user password following the procedure as described in steps 3 through 13 above Note This is the same user password as in the CALIBRAT mode Step 4 Press the...

Страница 16: ...40 1000 250 1600 400 2000 600 2500 1000 4000 5000 6000 10 000 psi psi psi bar kb cm2 bar Gauge Compound Absolute Gauge Compound vac 15 15 15 1 1 0 5 15 30 25 1 6 1 1 10 50 2 5 1 2 15 4 1 3 30 6 1 30 6...

Страница 17: ...00 9 00 Ranges Vac thru 7000 psi see engineering units below for other units of measurement Operating Temperature 0 150 F Storage Temperature 40 180 F DISPLAY Type LCD Display Digits 5 Character Heigh...

Страница 18: ...ns On Off 10 Sec 30 Sec 5 Min 15 Min Field Recalibration Zero Span password protected Contrast 7 Available Options Disable Config Options Allows disabling of Config Options password protected Calibrat...

Страница 19: ...Installation and Replacement The gauge comes standard with three AAA alkaline batteries installed For battery replacement use only one of the three types listed below Energizer AAA alkaline E92 LR03 A...

Страница 20: ...Ashcroft Inc 250 Main Street Stratford CT 06614 5145 Tel 203 378 8281 Fax 203 385 0602 e mail info ashcroft com www ashcroft com Ashcroft Inc 2013 I M008 10080 10 13 Rev E GBE...

Отзывы: