– 18 –
Enter
Selects Items In CONFIG Menu
Config Mode
Allows Scrolling Through CONFIG Menus
Engineering Units
Psi, ˝Hg, ˝H
2
O (with three temperature options, 60°F, 4°C
and 20°C), ft.SW, bar, mbar, kPa, mPa, mmHg, cmH
2
O,
mmH
2
O, kg/cm
2
Update Rate
4 Options: 10x/sec, 5x/sec, 2x/sec, 1x/sec
Auto Off
6 Options: Never, 2 Min., 5 Min., 15 Min., 30 Min.,
Dampening
6 Options: None, Average 2, 4, 6, 8 x update rate
Languages
7 Languages: English, Spanish, French, Italian, German,
Portuguese, Dutch
Backlite
5 Options: On/Off, 10 Sec., 30 Sec., 5 Min., 15 Min.
Field Recalibration
Zero & Span (password protected)
Contrast
7 Available Options
Disable Config Options
Allows disabling of Config Options (password protected)
Calibration Chart
10 Point Individual NIST Traceable Calibration Chart,
Standard
Standard Accessories
300 Series SS Protective Cover
Nylon Protective Carrying Pouch
Display
Description
Flashing
Gauge over/underpressured
0% or 100%
beyond 105% of range
LOW BAT
Low battery, replace
ERROR
Internal error, call customer
service (203) 378-8281
Pressure unit conversion exceeds
RES ERROR
display resolution or gauge
pressured beyond resolution
WARNING AND ERROR MESSAGES
SPECIFICATIONS