background image

ASG, Division of Jergens, Inc.

15700

 S. Waterloo Road  |  Cleveland, OH 44110-3898  

Phone: (888) 486-6163  |  Fax: (216) 481-4519
Email: [email protected]  |  Web: www.asg-jergens.com

Page 1 of 15

© 2017 Jergens, Inc.  All Rights Reserved   |   Revision Date: 02/19/18

DTT-L Series Digital Torque Testers

 

User Manual

Содержание DTT-L 10

Страница 1: ...oo Road Cleveland OH 44110 3898 Phone 888 486 6163 Fax 216 481 4519 Email asginfo asg jergens com Web www asg jergens com Page 1 of 15 2017 Jergens Inc All Rights Reserved Revision Date 02 19 18 DTT L...

Страница 2: ...ver Adapter 6 Usage 7 Procedure for Manual Torque Screwdrivers 7 Procedure for Power Screwdrivers 7 Battery Indicator 8 Basic Function 8 Display of Clockwise Counter Clockwise 8 Zeroing the Tester 9 C...

Страница 3: ...d or implied and all implied warranties of merchantability and fitness for a particular purpose are disclaimed and excluded from this transaction and shall not apply to the goods sold hereunder The AS...

Страница 4: ...ble 9 Volt 670 mA 90 264 VDC ASG 66619 ADC 16 Bit CPU 8 Bit Key Button 8 Buttons Unit of Measure 6 7 User Selectable Torque Units Depending on Model Mode of Measure 3 Modes Track Peak and First Peak D...

Страница 5: ...eters are supplied with the following standard items DTT L torque meter Battery charger Power adapter NOTE Rundown adapters sold separately International charger also sold separately DTT L Series Digi...

Страница 6: ...4 5 26 0 0 51 2 94 66618 1 4 Hex 25 0 80 0 2 8 9 03 66639 5 mm Hex 25 0 80 0 2 8 9 03 66634 3 8 Hex 50 0 200 0 5 65 22 6 66642 3 8 Hex 50 0 500 0 5 65 56 5 Below is a list of all rundown adapters ASG...

Страница 7: ...to be sure the readings are consistent The proper reading may be an average of several readings Procedure for Power Screwdrivers Put the appropriate rundown adapter into the unit s female 3 8 square...

Страница 8: ...tery Level 4 65V If the battery level is less than 4 6 V the Battery Empty message will be displayed and the tester will power down automatically NOTE Only use the 9V adapter charger supplied Clockwis...

Страница 9: ...m lbf in and lbf ft To change the display units press the UNIT key Each time the UNIT key is pressed it will select the next available unit until the tester returns to its original setting The DTT L...

Страница 10: ...ROW keys to move the cursor to SETTING Press ENTER To move between the submenus listed in the SETTING menu press the UP or DOWN ARROW keys to move the cursor The sub menus in the setting menu are auto...

Страница 11: ...menu page Use the UP and DOWNARROW keys to change the value Use the RIGHTARROW key to change the unit Press ZERO key to reset value to zero Press ENTER key to set new value and return to setting menu...

Страница 12: ...To access the Pass Fail menu press the MENU ESC key the display will show the main menu page Use the UP and DOWN ARROW keys to move the cursor point to SETTING Press the ENTER key and use the UP and D...

Страница 13: ...ess the MENU ESC key the display will show the main menu page Use the UP and DOWN ARROW keys to move the cursor point to MEMORY Press the ENTER key Press the UP and DOWN ARROW keys to move the cursor...

Страница 14: ...key the meter will return to the memory menu page If you select YES and press the ENTER key the meter will delete all saved records and return to the memory menu page DTT L Series Digital Torque Teste...

Страница 15: ...o access the Transducer State menu press the MENU ESC key the display will show main menu page Use the UP and DOWN ARROW keys to move the cursor point to TRANSDUCER STATE Press the ENTER key Press the...

Отзывы: