7

To operate, remove the saliva ejector valve

from the holder and insert a standard

saliva ejector tip.  Turn the vacuum toggle

switch, located on the right side of the cart,

to the 'ON' position. The vacuum may be

adjusted by moving the lever valve on the

saliva ejector tip. The lever valve assembly

snaps apart at the swivel and may be

autoclaved.
Waste from the vacuum system will collect

in the attached plastic bottle. When the

waste container is 3/4 full, it must be

emptied to prevent a back-up in the

system. After each use, the vacuum

system should be cleaned by running

clean water or vacuum system cleanser

through the evacuator hose and thoroughly

cleaning the waste bottle. 

CAUTION:

Use

only NON-foaming cleansers in the

vaccum lines. 

10. AA-290CK CENTRAL VACUUM KIT

(Optional) - 

The AA-290CK Kit (Fig.13) is

used with a central vacuum system. It

includes a central vacuum canister with

disposable solids collector, an autoclavable

central vacuum lever valve (with 5 feet of

tubing), and an autoclavable saliva ejector

tip valve (with 5 feet of tubing). An

industry-standard 1/2-inch connection hose

provides easy adaptation to the central

vacuum system.

WASTE

CONTAINER

VACUUM HOSE

QUICK DISCON-

NECT

Figure 12 - Optional AA-25 LVE

Vacuum System

Figure 13 - Optional AA-290CK

Central Vacuum Kit

SALIVA EJECTOR

VALVE W/LEVER

WASTE CONTAINER

QUICK DISCONNECT

Содержание ADU-25

Страница 1: ...OPERATION and MAINTENANCE INSTRUCTION MANUAL ADU 25 Custom Cart...

Страница 2: ...7 Handpiece Controls 4 Fig 8 Foot Control 4 Fig 9 Water Tank Removal 5 Fig 10 Right Side of Chassis 6 Fig 11 AA 21 HVE Vacuum Sys Opt 6 Fig 12 AA 25 LVE Vacuum Sys Opt 7 Fig 13 AA 290CK Central Vac K...

Страница 3: ...ions provided in this manual to receive the best and longest service from your Aseptico equipment Separate manuals may be provided to cover the operation and maintenance of hand held pieces or other a...

Страница 4: ...E VACUUM SYSTEM Optional THREE WAY SYRINGE NWS 8 HDPE WATER SUPPLY TANK Optional HIGHSPEED HANDPIECE CONTROL AA 25 SALIVA EJECTOR SYSTEM Optional REMOVABLE INSTRUMENT TRAY 3 The ADU 25 unit must be co...

Страница 5: ...connect at the rear of the delivery head Fig 4a Note manual release lever on disconnect Insert metal fitting on top of bottle into quick disconnect Locking ring will snap into place Press release leve...

Страница 6: ...e locking collar surrounding the tip socket and pull the used tip straight out of the socket see Fig 6 To insert a new tip press locking collar and push tip into socket as far as it will go Release co...

Страница 7: ...ol 7 FLUSH TOGGLE Fig 10 The flush toggle located on the right side of the chassis allows you to quickly and completely flush your handpieces washing away contaminants which may have accumulated in th...

Страница 8: ...The ADU 25 is supplied with a large capacity container 9 AA 25 SALIVA EJECTOR SYSTEM Optional The AA 25 Saliva Ejector is an air powered self contained oral evacuation system It is supplied with a pla...

Страница 9: ...clean water or vacuum system cleanser through the evacuator hose and thoroughly cleaning the waste bottle CAUTION Use only NON foaming cleansers in the vaccum lines 10 AA 290CK CENTRAL VACUUM KIT Opti...

Страница 10: ...erclockwise will increase pressure 3 HANDPIECE PRESSURE ADJUSTMENT Fig 15 The amount of air pressure supplied to each handpiece may be adjusted by turning the flow control for that handpiece on the co...

Страница 11: ...onds after every patient and about 20 seconds at the beginning of each day Refer to page 4 3 GENERAL CLEANING The cart exterior is designed to make cleaning and the use of prophylactic barriers as eas...

Страница 12: ...rt or whole Aseptico shall be the sole arbiter of such action In the event of alleged defect under warranty the purchaser is to notify Aseptico s Customer Service Department promptly Customer Service...

Отзывы: