![ASD FieldSpec 3 Скачать руководство пользователя страница 81](http://html1.mh-extra.com/html/asd/fieldspec-3/fieldspec-3_user-manual_2981633081.webp)
ASD Document 600540 Rev. J
71
FieldSpec® 3 User Manual
www.asdi.com
Chapter 7 Field Measurements
Step 2
Place the glass into the sink.
Step 3
Place 220 grade wet sandpaper onto the glass.
Step 4
Gently move the Spectralon reference in a figure 8 motion on the sandpaper,
using water as needed to wash away the thin layer that is sanded off.
Step 5
Blow dry with clean air or nitrogen or allow the material to air dry.
Step 6
If the material requires high resistance to deep UV radiation, the piece should
be prepared as above, then either of the following two treatments performed.
1
Flush the Spectralon piece with >18 milli-ohm distilled, deionized water
for 24 hours.
2
Vacuum bake the Spectralon piece at 75° C for a 12 hour period at a
vacuum of 1 Torr or less. Then purge the vacuum oven with clean dry air
or nitrogen.
WARNING!
Do not use oils or soaps to clean the Spectralon white
reference.
7.11.3
White Reference Procedures
White references should be collected approximately every 15 minutes and can
be varied depending on whether conditions are changing rapidly or not
changing very much. The Spectralon puck should be used when optimizing
and taking a white reference measurement.
Note:
For best results, the Spectralon puck should be at the same distance and angle
from the foreoptic as the sample will be during measurements.
When saving reflectance data, point the probe at the Spectralon with the same
viewing geometry for a minute or two every few measurements. If the relative
reflectance of the Spectralon is less than or greater than one, a new white
reference may be needed. If the relative reflectance of the Spectralon is
greater than one, re-optimization is recommended.
The FieldSpec spectroradiometer should be re-optimized for:
•
Light changes.
•
Any atmospheric changes.
•
Temperature changes.
•
Whenever accessory probes are changed.
Note:
Environmental conditions can change rapidly or slowly. It all depends on
clouds, wind (affecting temperature), instrument warm up time, etc.
Содержание FieldSpec 3
Страница 1: ...FieldSpec 3 User Manual ASD Document 600540 Rev J 2010 by ASD Inc www asdi com...
Страница 7: ...ASD Document 600540 Rev J vii FieldSpec 3 User Manual www asdi com...
Страница 10: ...ASD Document 600540 Rev J x FieldSpec 3 User Manual www asdi com Notes...
Страница 46: ...ASD Document 600540 Rev J 36 FieldSpec 3 User Manual www asdi com Chapter 4 Fiberoptic Interface Notes...
Страница 50: ...ASD Document 600540 Rev J 40 FieldSpec 3 User Manual www asdi com Chapter 5 Batteries Notes...
Страница 82: ...ASD Document 600540 Rev J 72 FieldSpec 3 User Manual www asdi com Chapter 7 Field Measurements Notes...
Страница 89: ...ASD Document 600540 Rev J 79 FieldSpec 3 User Manual www asdi com Appendix A Troubleshooting Notes...
Страница 90: ...ASD Document 600540 Rev J 80 FieldSpec 3 User Manual www asdi com Appendix A Troubleshooting...
Страница 108: ...ASD Document 600540 Rev J 96 FieldSpec 3 User Manual www asdi com Appendix B Reference Information Notes...