ASD Document 600540 Rev. J
67
FieldSpec® 3 User Manual
www.asdi.com
Chapter 7 Field Measurements
Spectralon is an example of a reference standard to take a white reference
against. (Labsphere, Inc., Shaker St., PO Box 70, North Sutton, NH
03260-0070, PH: 603-927-4266, FX: 603-927-4694, E-mail:
[email protected], Website: www.labsphere.com). Spectralon is also
available from ASD.
Step 6
Take a
White Reference
(or baseline reference).
Step 7
Observe samples.
7.10.2 Indico Application
Step 1
Follow the instructions for plugging in the instrument and starting the Indico
software.
Step 2
Attach any required foreoptic accessories.
Step 3
Open a new project either by selecting
File->New Project
from the
pull-down menu.
Step 4
Type in a name for this project and click on the
OK
button.
Step 5
Select
Spectrum->Sample Count/Average
from the pull-down menu.
Step 6
Set the
Instrument Sample Count
to 100 for outdoor solar illumination
and click
OK
.
Note:
The actual spectrum average will be determined by striking a
compromise between noise reduction through averaging the spectra
and the time desired for each spectrum collection. For instance, if you
are using the instrument in the field, are walking a large area, and are
making frequent spectral readings, you will want a shorter average
setting than if you are collecting spectra in-situ and desire the cleanest
spectra possible.
The project screen will return, and the sample count will be displayed in the
status box at the bottom of the graph.
Step 7
Make sure the light source is on and the probe input end is pointed at the
white reference panel.
Step 8
Select
Spectrum->Take a Baseline [Alt+S, B] [Ctrl+R]
. When
prompted, click
Yes, take a new white reference
.
•
The FieldSpec spectroradiometer optimizes the detector sensitivities for
the probe and light source currently being used.
•
The dark offset and white reference will also be measured and saved.
•
Status bars will indicate each process.
When optimization and white reference are complete, you should see a
straight baseline across the Project Graph at 1.00 (100%).
Step 9
Select the
Display->Reflectance [Alt+D, R]
for the reflectance mode.
Содержание FieldSpec 3
Страница 1: ...FieldSpec 3 User Manual ASD Document 600540 Rev J 2010 by ASD Inc www asdi com...
Страница 7: ...ASD Document 600540 Rev J vii FieldSpec 3 User Manual www asdi com...
Страница 10: ...ASD Document 600540 Rev J x FieldSpec 3 User Manual www asdi com Notes...
Страница 46: ...ASD Document 600540 Rev J 36 FieldSpec 3 User Manual www asdi com Chapter 4 Fiberoptic Interface Notes...
Страница 50: ...ASD Document 600540 Rev J 40 FieldSpec 3 User Manual www asdi com Chapter 5 Batteries Notes...
Страница 82: ...ASD Document 600540 Rev J 72 FieldSpec 3 User Manual www asdi com Chapter 7 Field Measurements Notes...
Страница 89: ...ASD Document 600540 Rev J 79 FieldSpec 3 User Manual www asdi com Appendix A Troubleshooting Notes...
Страница 90: ...ASD Document 600540 Rev J 80 FieldSpec 3 User Manual www asdi com Appendix A Troubleshooting...
Страница 108: ...ASD Document 600540 Rev J 96 FieldSpec 3 User Manual www asdi com Appendix B Reference Information Notes...