Before Starting Centrifuge 

The main switch is located on the lower right side of the centrifuge.

Before starting up the centrifuge for the first time, make sure that the rotor locking
knob is securely fastened.

Safety Precautions 

 Always 

The rotor and lid must 

always

 be securely fastened.

If the centrifuge makes unusual noises

 when started, or exhibits excessive

vibration, switch the unit off 

immediately

 by turning the power switch OFF.

The centrifuge

 

must be loaded symmetrically

.

Do not use centrifuge with wet hands.

Take all precautions when using the centrifuge for toxic, radioactive, or material

with pathogenic micro-organisms, with respect to your safety and the safety of
others.

Do not place any sample containers or test tubes directly into the rotor.

 

The Centrifuge Itself 

Do not use a centrifuge that was not correctly installed or repaired.

This centrifuge may be used only for the specified application.  It must not be

operated in a hazardous or flammable environment and must not be used to
centrifuge flammable, explosive, or highly reactive substances.

When moving a centrifuge from a cold room to one of normal temperature, run

the centrifuge for 30 minutes in the cold room so that it does not get covered in
condensation.  Alternatively allow it to warm up for at least three hours,
unplugged, in order to prevent condensation.

The rotor and polycarbonate carrier are high-grade components which are

subject to extreme mechanical stress.  Slight scratches, tears, dents, or other
damage can lead to failure of the centrifuge.

All material associated with the centrifuge should be protected from excessive

ultra-violet radiation.

Reminders 

If any function of the centrifuge does not operate correctly do not use the

centrifuge.

Do not use a damaged centrifuge.

Do not move the centrifuge during the run.

After prolonged use, or if the motor was stalled, the bottom of the centrifuge may

become very hot.  Handle the centrifuge by the sides only.

Содержание CYGNUS CF VI

Страница 1: ...1 CYGNUS CF VI BLOOD CENTRIFUGE OPERATORS MANUAL Revised December 2011 ...

Страница 2: ...entrifuge Safety Precautions Operation Opening Balancing Loading Starting Verifying Speed Cleaning Centrifuge Rotor Carrier Maintenance Inspection Contact Information for Questions Assistance Centrifuge Replacement Before starting this centrifuge for the first time please read this short manual ...

Страница 3: ...evel of transient over voltage is Category II Rotating speed 3450 100 RPM Items included Operators Manual Centrifuge Power Cord Safety Shields polycarbonate carrier Each centrifuge comes with six empty polycarbonate carriers Installing Centrifuge Place the centrifuge on a level horizontal surface Make sure that the ventilation slots on the bottom are not blocked in accordance with safety regulatio...

Страница 4: ... be used only for the specified application It must not be operated in a hazardous or flammable environment and must not be used to centrifuge flammable explosive or highly reactive substances When moving a centrifuge from a cold room to one of normal temperature run the centrifuge for 30 minutes in the cold room so that it does not get covered in condensation Alternatively allow it to warm up for...

Страница 5: ...carriers with blood collection tubes into rotor Load must be balanced at all times by placing carriers in the rotor in a balanced configuration There must be at least two 2 carriers placed opposite each other Three 3 or four 4 carriers must be evenly spaced Five 5 carriers cannot be balanced Each carrier must contain equal weight Improper balancing usually causes excessive vibration and will shut ...

Страница 6: ...detergent Spills should be wiped up immediately Clean both the exterior and the interior Interior cleaning includes the interior bucket specimen holder and rubber cushion Disinfect using a 10 bleach solution one part bleach to nine parts water Exterior of the centrifuge should be disinfected weekly Interior should be disinfected monthly Prepare a fresh 10 bleach solution each time Once mixed with ...

Страница 7: ...ed insulation broken pins any damage Notes o Since carriers are used there are no bioshields or biosafety components o The carrier may be autoclaved o There are no user serviceable parts in the centrifuge This centrifuge uses a non replaceable rotor Repairs must be done by an authorized service agency Contact Information for Questions Assistance Centrifuge Replacement Ascend Clinical Client Servic...