Introduction
This is a bench top personal centrifuge with a space saving design and intuitive controls.
The unit has provision for six tubes in the rotor, cycle timer, automatic lid lock and
release.
It automatically shuts off after the 20 minute cycle is complete, and the lid automatically
opens after the centrifuge has come to a complete stop.
Equipment Ratings, Speed & Items Included
This centrifuge is for indoor use only
Maximum altitude of operation: 9,000 feet
Maximum relative humidity of 80% for temperatures up to 31
C decreasing
linearly to 50% relative humidity at 40
C
Transient over voltages on the main power supply. The normal level of transient
over voltage is Category II.
Rotating speed: 3450±100 RPM
Items included
Operators Manual
Centrifuge
Power Cord
Safety Shields (polycarbonate carrier)
Each centrifuge comes with six empty polycarbonate carriers.
Installing Centrifuge
Place the centrifuge on a level, horizontal surface.
Make sure that the ventilation slots on the bottom are not blocked in accordance
with safety regulations of IEC 1010-2-020 (equivalent to EN 61010-2-020).
A safety distance of 12 inches is needed around the centrifuge during operation.
No objects which could cause additional damage, or could themselves be
damaged in the event of a centrifuge crash should be positioned in this space.
Before plugging in the centrifuge compare your main supply with the electrical
requirements listed on the identification plate on the lower portion of the unit.