Electromagnetic Emissions
NanoScope Manual
DFU-0298-1, Revision 0
Page 38 of 38
Table 14
Guidance and Manufacturer's Statement – Recommended Separation
Recommended separation distances between portable and mobile RF
communications equipment and the NanoScope System.
Rated maximum output
power of transmitter [W]
Separation distance according to frequency of transmitter [m]
150 kHz to 80
MHz
d = 1.2 √P
80 MHz to
800 MHz
d = 1.2 √P
800 MHz to
2.7 GHz
d = 2.3 √P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance
d
in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where
P
is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Содержание NanoScope
Страница 43: ...This page intentionally left blank ...