Electromagnetic Emissions
NanoScope Manual
DFU-0298-1, Revision 0
Page 37 of 38
Table 13 (cont.)
Guidance and Manufacturer's Statement - Electromagnetic
Immunity
The NanoScope System is intended for use in the electromagnetic environment
specified below. The customer or the user of the NanoScope System should ensure
that it is used in such an environment.
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to
80 MHz
10 V/m
80 MHz to
2.7 GHz
3 Vrms
10 V/m
Portable and mobile RF communications
equipment should be used no closer to any
part of the NanoScope System, including
cables, than the recommended separation
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance:
[
]
P
V
d
1
/
5
.
3
=
= 1.2
P
[
]
P
V
d
1
/
5
.
3
=
=1.2
P
80 MHz to
800 MHz
[
]
P
E
d
1
/
7
=
= 2.3
P
800 MHz to 2.7 GHz
Where
P
is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and
d
is the
recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
1
should be less than the compliance level in
each frequency range.
2
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.