INSTRUCTIONS FOR USE
Document Number: D-720763
Page 23
Issue Date: 31 MAR 2020
Version: C
Ref Blank Template: 80025118 Ver. E
top of the Patient Platform on the operative side, pivot the Post into the
opening on the Patient Platform and release the collar (Fig 13C). Ensure that
the collar engages the opening in the Patient Platform and locks the Post in
place by pulling on the post to ensure engagement.
Place the reusable Perineal Post Pad (AR-6529-08 or AR-6529-21) onto the
Perineal Post. Place the Disposable Post Pad Cover over the Post following the
directions for the Cover. Rotate so that the seam faces away from the patient.
Positioning the Patient:
Unlock the Carriages on the Well Leg and Operative Spars.
Slide the patient distally, so that the patient’s perineum is positioned against
the Perineal Pad. While the patient is being slid distally, a member of the staff
can also pull the carriages on the Well Leg and Operative Spars distally so that
the patient’s legs are fully extended.
Positioning the Spars:
Abduction / Adduction
Abduction and/or Adduction of the patient’s leg may
be achieved by actuating the squeeze trigger on the
Spar handle (Fig 14A). Actuating this trigger will permit
the Spar to be adducted (pivoted towards the mid-line
of the patient) or abducted (pivoted away the mid-
line of the patient). This feature is available on both the
Operative and Well Leg Spars (Fig 14B).
Figure 14
B
A
A
B
C
Mount
Figure 13
Содержание AR-6529S
Страница 1: ...Hip Distraction System Instructions for Use Product No AR 6529S D 720763 Version C ...
Страница 34: ...Hip Distraction System Instructions d utilisation N de produit AR 6529S D 720763 Version C ...
Страница 67: ...Hip Distraction System Gebrauchsanleitung Produkt Nr AR 6529S D 720763 Version C ...
Страница 101: ...Hip Distraction System Istruzioni per l uso N prodotto AR 6529S D 720763 Version C ...
Страница 135: ...Hip Distraction System Instrucciones de uso N º de producto AR 6529S D 720763 Version C ...
Страница 173: ......
Страница 174: ......