INSTRUCTIONS FOR USE
Document Number: D-720763
Page 18
Issue Date: 31 MAR 2020
Version: C
Ref Blank Template: 80025118 Ver. E
Fig: 6
c.
Slide carriage so that it is located about 10 in (25.5cm) from the Spar joint and
pull the Carriage Lock lever to lock the Carriage in place.
d.
Ensure that the Mounting Hook located on the underside of the Spar joint is in
line with long axis of spar. If necessary, adjust by actuating the Abduction /
Adduction handle (Fig 7) and having a co-worker adjust the angle of the
mounting hook so that it is in line with the axis of the spar.
Fig: 7
e.
Lift the Spar and advance at about a 30
⁰
angle to engage the Mounting Hook
with the Mounting Pin (Fig 8A) on the operative side of the Patient Platform.
Lower the Spar (Fig 8B) to fully mount the Operative Spar to the Patient
Platform. Secure the Spar in place by tightening the Locking Knob (Fig 8C) on
the lateral side of the Patient Platform so that the red O-rings are no longer
visible.
Caution: Ensure that the red O-rings are no longer visible to ensure the knob is
fully tightened.
AB/Adduction
trigger
Содержание AR-6529S
Страница 1: ...Hip Distraction System Instructions for Use Product No AR 6529S D 720763 Version C ...
Страница 34: ...Hip Distraction System Instructions d utilisation N de produit AR 6529S D 720763 Version C ...
Страница 67: ...Hip Distraction System Gebrauchsanleitung Produkt Nr AR 6529S D 720763 Version C ...
Страница 101: ...Hip Distraction System Istruzioni per l uso N prodotto AR 6529S D 720763 Version C ...
Страница 135: ...Hip Distraction System Instrucciones de uso N º de producto AR 6529S D 720763 Version C ...
Страница 173: ......
Страница 174: ......