685 Series
ComboPak
User’s Manua
l · Page 13
Menu
Description
Factory
Default
Computer
Command
Ext Fan
This selects the voltage for the
auxiliary fan power supply. See the
External Fan Control
section
below
for more information.
Off
TEC:FAN
Ext Fan Pwr
When
Ext Fan
is set to
Custom
, this
sets the specific fan voltage from
4.0 to 12.0 volts. See the
External
Fan Control
section
below for more
information.
12.0
TEC:FAN
Ext Fan
Mode
Controls when the fan operates.
See the
External Fan Control
section
below for more information.
Auto
TEC:FAN
Ext Fan Off
When
Ext Fan Mode
is set to
Delay
, this setting defines the
number of minutes to delay turning
off the fan after the TEC output has
been turned off. See the
External
Fan Control
section
below for more
information.
5 minutes
TEC:FAN
Cable R
The resistance of the cable and
connectors, in ohms. This setting
allows for accurate voltage
measurement at the TEC by
removing the voltage loss of the
cable.
0.0080Ω
TEC:CABLER
T Rate
Selects a desired temperature ramp
rate in degrees Celsius per minute.
Set to 0.0°C/min to disable rate
limiting.
0.0°C/min
TEC:TRATE
TEC Sensor Settings
Thermistor
coefficients
Thermistor coefficients in the
Steinhart-Hart equation.
A= 1.12924E-03
B= 2.34108E-04
C= 0.87755E-07
TEC:CONST
Communications Settings
Err While
Rmt
To turn off the display of errors
while in remote mode, set this value
to “No”. To display errors while in
remote mode, set this value to
“Yes”.
Yes
REMERR