Page 12 · 685 Series
ComboPak
User’s Manua
l
Menu
Description
Factory
Default
Computer
Command
T-High Lim
T-High Lim
is the upper
temperature limit. If operating in T
Mode and the actual temperature
rises above this value, the output
will be turned off. Temperature
limits not monitored in R Mode or
Ite Mode. Limit monitoring can also
be disabled in software (see
TEC:ENAB:OUTOFF command).
125°C
TEC:LIM:THI
R-Low Lim
R-Low Lim
is the sensor low limit. If
operating in R Mode and the actual
sensor measurement drops below
this value, the output will be turned
off. R limits are not monitored in Ite
Mode. Limit monitoring can also be
disabled in software (see
TEC:ENAB:OUTOFF command).
0.01kΩ
TEC:LIM:RLO
R-High Lim
R-High Lim
is the sensor high limit.
If operating in R Mode and the
actual sensor measurement
exceeds this value, the output will
be turned off. R limits are not
monitored in Ite Mode. Limit
monitoring can also be disabled in
software (see TEC:ENAB:OUTOFF
command).
45kΩ
TEC:LIM:RHI
Tol Time
Tolerance time is the amount of
time, in seconds, that the actual
temperature must be within the set
point tempe/- the Tol Temp
value for the unit to be considered
in tolerance.
5 seconds
TEC:TOL
Tol Temp
Tolerance temperature is a
temperature band (in °C) around
the set point temperature. When the
actual temperature is within this
band for longer than the Tol Time
setting, then the unit is considered
to be in tolerance.
0.1°C
TEC:TOL
H/C Mode
This selects the heating and/or
cooling mode of the
ComboPak
.
See the
Resistive Heaters and
Heat/Cool Only Modes
, below, for
more information.
Heat/Cool
TEC:HEATCO
OL