Armstrong Steam QM-3 Скачать руководство пользователя страница 3

Armstrong International
IOM-246-V2.0
Steam QM-3 Calibration

Page 1 of 14

 

Revision History 

Version

release Date

Description of Changes

1.0

20/02/2014

Initial

1.1

21/03/2014

Temperature transmitter T (T1, T2, T3) - Design Modification

2.0

17/03/2016

Tee and Adapter - Design Modification: pages 5, 6, 8, 9.

Page 4, General Information Section.  Update of the list of material needed

The following have been removed: Decade box (calibrated), 5 volt generator or adjustable power 

supply (0-10VDC), M12 digital for decade box

The following have been added: Oil bath, Thermometer (calibrated), Pressure generator, 

Manometer (calibrated).  

Page 6, Recalibration Procedure Section

Point 3 has been added (Remove the Temperature Sensor T1 after unplugging it and plug it back 

in before install it in oil bath).

- "Calibrating the Temperature Sensors (Simulated Measurement Method)" has been updated as 

"Calibrating the Temperature Transmitter".

- Chart with the data of the 3 temperature sensors is deleted.

Pages 6 - 8: Calibrating the temperature transmitter and the pressure transmitter procedures have 

been updated.

Page 12: Text modifications, Point 1(Disconnect the expansion coil from the bottom of the NCG 

burette. To avoid damaging the expansion coil, loosen bottom nut on expansion coil and remove 

top nut). Point 13 (o force open the EV1 valve to allow discharge move the arrow to the EV1 line 

then move the arrow to left of 1 and press OK).

Содержание Steam QM-3

Страница 1: ...auts Sarts 2 me avenue 4 4040 HERSTAL BELGIUM T l 32 0 4 240 90 90 Fax 32 0 4 240 40 33 armstronginternational com Keep this manual with equipment for future reference STEAM QM 3 Steam Quality Monitor...

Страница 2: ...ginternational com Armstrong Contents Revision History 1 Safety 2 Abbreviations and Acronyms 3 General Information 4 Recalibration Procedure 6 Calibrating the Temperature Sensors 6 Calibrating the Pre...

Страница 3: ...librated Page 6 Recalibration Procedure Section Point 3 has been added Remove the Temperature Sensor T1 after unplugging it and plug it back in before install it in oil bath Calibrating the Temperatur...

Страница 4: ...ualified person Installation should always be accompanied by competent technical assistance Improper installation start up operation maintenance or service may void warranty You are encouraged to cont...

Страница 5: ...seconds DIN Deutsches Institut f r Normung eV dP Differential Pressure EC European Community EEC European Electrotechnical Commission EN European Norm EV Electronic Valve F Fahrenheit gal Gallon h Hou...

Страница 6: ...nsors in the oil bath and recreate the actual temperatures for calibration This method is not covered in this document Pressure transmitter The pressure sensor is calibrated by simulating the voltage...

Страница 7: ...ibrated on the Steam QM 3 The instruments are 3 temperature sensors 1 pressure transmitter 1 heating element 1 condensate flow meter 1 gas flow meter NCG meter 2 differential pressure sensors Heating...

Страница 8: ...4 If not you must adjust the sensor 11 On the Steam QM 3 display pad press left and right arrows simultaneously to access the code menu 12 Enter 234 using the up and down arrows to change numbers and...

Страница 9: ...edure in step 1 for T3 Temperature sensor NOTE Any corrections to the calibration certificate must be done following good documentation practices The error is crossed through by a single line The corr...

Страница 10: ...the Steam QM 3 display pad press up and down arrows simultaneously to access the Debug sensor menu 7 Write the value of P1 on the calibration certificate If P1 is 0 1 bar regarding the set point 0 bar...

Страница 11: ...d press left and right arrows simultaneously to access the code menu 7 To access the resistance calibration menu enter 069 using the up and down arrows to change numbers and the left and right arrows...

Страница 12: ...alve from the draw vessel slowly to establish a cycle rate of the EV2 valve of approximately 30 seconds One cycle is from opening to closure of valve you should be able to hear this If the environment...

Страница 13: ...ertificate NOTE Any corrections to the calibration certificate must be done following good documentation practices The error is crossed through by a single line The correction is written alongside wit...

Страница 14: ...r syringe before resetting to 2ml 8 With the NCG injector syringe disconnected reset the syringe to 2ml 9 Connect the NCG injector syringe to the middle of the NCG injector assembly tee with a 6mm tub...

Страница 15: ...w sensor calibration menu 24 On the Steam QM 3 display pad press left and right arrows simultaneously to access the code menu 25 To access the flow sensor calibration menu enter code 123 using the up...

Страница 16: ...er draw vessel connecting to the bottom of the NCG burette use two wrenches 10mm and 11mm to ensure no damage to the NCG burette occurs Reconnect the expansion coil to the bottom of the NCG burette us...

Отзывы: