Steam QM-3 Calibration

Page 12 of 14

 

Calibration #7: Calibrating the nCG Flow Meter

  

  1.   Disconnect the expansion coil from the bottom of the NCG burette. To avoid damaging the expansion coil Loosen  

 

        bottom nut on expansion coil and remove top nut.—use two wrenches 10mm and 11mm to 

 

      ensure 

no

 damage to the NCG burette occurs—and swing out of the way.

 

2.  Place the 1 liter draw vessel filled with tap water on top of the Steam QM-3 so as to have good gravity flow.

 

3.  Connect the 6mm tube coming from the 1 liter draw vessel to the bottom of the NCG burette (where the expansion.   

 

    coil was connected—use two wrenches 10mm and 11mm to ensure no damage to the NCG burette occurs).                                                                            

 

4.  Disconnect the 6mm tube coming from P1 where it connects to the top of the NCG burette. 

 

5.  Connect the NCG injector assembly tee to the top of NCG burette. 

 

      6.   Connect the 6mm tube that was disconnected from the top of the NCG burette to the top of injector tee. 

 

      7.   To avoid damage to P1 by creating a vacuum, always remove the NCG injector syringe before resetting   

 

       to 

2ml.

 

      8.   With the NCG injector syringe disconnected, reset the syringe to 2ml.

 

      9.   Connect the NCG injector syringe to the middle of the NCG injector assembly tee with a 6mm tube.

 

10.  Force open the EV1 valve using the pressure calibration menu 152.

 

11.  On the Steam QM-3  display  pad, press left and right arrows simultaneously to  access the  code  menu.                                                    

 

12.  To access the pressure calibration menu, enter code 152 using the up and down arrows to change numbers, and the  

 

    left and right arrows to change the position.

  

Calibration

P Sensor

P1

0 barg

3 barg

EV1

0

1

115

230

Pressure calibration menu 152

 

13.  To force open the EV1 valve to allow discharge move the arrow to the EV1 line then move the arrow to left of 1 and  

 

    press OK.

 

14.  Open the valve on the 1 liter draw vessel to fill NCG burette and allow water to run into the waste container.

 

15.  Close the valve on the 1 litter draw vessel.

 

16.  Close the EV1 valve using menu 152 on the control panel, move the arrow to left of 0 and press OK.

 

    17.  To avoid damage to P1 by creating a vacuum, always remove the NCG injector syringe before resetting to 2ml.

        

 

 

 

Calibrating the nCG Flow Meter 

'

3

Expansion coil swung out of the way

6 mm tube

1

EV1

syringe

valve

valve

1 liter draw vessel

Содержание Steam QM-3

Страница 1: ...auts Sarts 2 me avenue 4 4040 HERSTAL BELGIUM T l 32 0 4 240 90 90 Fax 32 0 4 240 40 33 armstronginternational com Keep this manual with equipment for future reference STEAM QM 3 Steam Quality Monitor...

Страница 2: ...ginternational com Armstrong Contents Revision History 1 Safety 2 Abbreviations and Acronyms 3 General Information 4 Recalibration Procedure 6 Calibrating the Temperature Sensors 6 Calibrating the Pre...

Страница 3: ...librated Page 6 Recalibration Procedure Section Point 3 has been added Remove the Temperature Sensor T1 after unplugging it and plug it back in before install it in oil bath Calibrating the Temperatur...

Страница 4: ...ualified person Installation should always be accompanied by competent technical assistance Improper installation start up operation maintenance or service may void warranty You are encouraged to cont...

Страница 5: ...seconds DIN Deutsches Institut f r Normung eV dP Differential Pressure EC European Community EEC European Electrotechnical Commission EN European Norm EV Electronic Valve F Fahrenheit gal Gallon h Hou...

Страница 6: ...nsors in the oil bath and recreate the actual temperatures for calibration This method is not covered in this document Pressure transmitter The pressure sensor is calibrated by simulating the voltage...

Страница 7: ...ibrated on the Steam QM 3 The instruments are 3 temperature sensors 1 pressure transmitter 1 heating element 1 condensate flow meter 1 gas flow meter NCG meter 2 differential pressure sensors Heating...

Страница 8: ...4 If not you must adjust the sensor 11 On the Steam QM 3 display pad press left and right arrows simultaneously to access the code menu 12 Enter 234 using the up and down arrows to change numbers and...

Страница 9: ...edure in step 1 for T3 Temperature sensor NOTE Any corrections to the calibration certificate must be done following good documentation practices The error is crossed through by a single line The corr...

Страница 10: ...the Steam QM 3 display pad press up and down arrows simultaneously to access the Debug sensor menu 7 Write the value of P1 on the calibration certificate If P1 is 0 1 bar regarding the set point 0 bar...

Страница 11: ...d press left and right arrows simultaneously to access the code menu 7 To access the resistance calibration menu enter 069 using the up and down arrows to change numbers and the left and right arrows...

Страница 12: ...alve from the draw vessel slowly to establish a cycle rate of the EV2 valve of approximately 30 seconds One cycle is from opening to closure of valve you should be able to hear this If the environment...

Страница 13: ...ertificate NOTE Any corrections to the calibration certificate must be done following good documentation practices The error is crossed through by a single line The correction is written alongside wit...

Страница 14: ...r syringe before resetting to 2ml 8 With the NCG injector syringe disconnected reset the syringe to 2ml 9 Connect the NCG injector syringe to the middle of the NCG injector assembly tee with a 6mm tub...

Страница 15: ...w sensor calibration menu 24 On the Steam QM 3 display pad press left and right arrows simultaneously to access the code menu 25 To access the flow sensor calibration menu enter code 123 using the up...

Страница 16: ...er draw vessel connecting to the bottom of the NCG burette use two wrenches 10mm and 11mm to ensure no damage to the NCG burette occurs Reconnect the expansion coil to the bottom of the NCG burette us...

Отзывы: