background image

 

18 

Routine Maintenance 

Responsibility 

To preserve the life and efficient operation of the equipment it is important that the 
equipment is properly maintained. Regular maintenance of the equipment is the 
responsibility of the end user and must be performed by qualified personnel who 
understand the operation of the equipment. 

General 

The equipment should be disconnected from the electrical supply when not in use. 
After use the feed bottles, reactor vessel, sump tray and pipework should be washed 
through with water to remove chemical residues and then drained.  

RCD Test 

Test the RCD by pressing the TEST button at least once a month. If the RCD button 
does not trip when the Test button is pressed then the equipment must not be used 
and should be checked by a competent electrician. 

Temperature sensor calibration 

The temperature sensors are calibrated before delivery and should not require re-
calibration. However should calibration become necessary follow the below 
procedure. This should only be done once the unit has fully warmed up. 

The temperature conditioning circuit (which provides the reading from the 
conductivity probe supplied with the CEXC service unit) is located on a printed circuit 
board inside the plinth on the right-hand side. This circuit is calibrated before 
despatch and should not require re-calibration. However, should re-calibration 
become necessary the appropriate calibration potentiometers can be located using 
the diagram given in the CEXC manual (Routine Maintenance). 

Connect CEXC service unit to a PC and start up the Armfield software. Open mimic 
diagram screen where T1, T2 and T3 windows are displayed. 

If a thermocouple calibrator is available: 

Connect Thermocouple calibrator simulator to T1 input socket. Set to 25ºC and 
adjust VR1 (T1 ZERO) and VR2 (T1 SPAN) on the PCB to give 25ºC displayed on 
PC. Check accuracy at 15º and 40ºC. 

Repeat the same procedure for T2 by adjusting VR3 (T2 ZERO) and VR4 (T2 SPAN) 
on the PCB to give 25ºC displayed on PC, and VR5 (T3 ZERO) and VR6 (T3 SPAN) 
for T3. 

If a thermocouple calibrator is not available: 

Temperature sensor T1, T2 and T3 should be dipped into crushed ice, and then 
adjust the ZEROS to give 0ºC, then sensors should be dipped into boiling water and 
then adjust the SPANS to 100ºC. 

Calibration of the conductivity sensor 

The Low and High conductivity conditioning circuits (which provide the reading from 
the conductivity probe supplied with the CEXC service unit) are located on the 
printed circuit board shown above inside plinth. These circuits are calibrated before 

Содержание CEM-MKII

Страница 1: ...Continuous Stirred Tank Reactor Instruction Manual CEM MKII ISSUE 20 September 2013...

Страница 2: ...Chemical Safety 6 Description 7 Overview 7 Flow of materials 7 Installation 9 Advisory 9 Installing the PC Software 9 Installing the Equipment 9 Operation 12 Operating the PC Software 12 Operating the...

Страница 3: ...mon Theory 22 Exercise A To find the reaction rate constant in a Continuous Stirred Tank Reactor 24 Exercise B To determine the effect of inadequate mixing on the reaction rate 29 Exercise C Determina...

Страница 4: ......

Страница 5: ...Customer Support helpdesk Monday to Thursday 0830 1730 and Friday 0830 1300 UK time Contact details are as follows United Kingdom International 0 1425 478781 calls charged at local rate 44 0 1425 4787...

Страница 6: ...th the CEXC Computer Controlled Service Unit An alternative instruction manual is available from Armfield that describes the use of the CEM MKII in conjunction with the CEX Service Unit Please contact...

Страница 7: ...rticularly suitable for liquid phase reactions It is particularly used in the organic chemicals industry Advantages include consistent product quality straightforward automatic control and low manpowe...

Страница 8: ...4 Equipment Diagrams Figure 1 Reactor Unit...

Страница 9: ...a will feed on any scale rust algae or sludge in water and will breed rapidly if the temperature of water is between 20 and 45 C Any water containing this bacterium which is sprayed or splashed creati...

Страница 10: ...ressed Failure to trip means that the operator is not protected and the equipment must be checked and repaired by a competent electrician before it is used Hot Surfaces and Liquids The unit incorporat...

Страница 11: ...ensors provided with the service unit The larger of the two glands is for the conductivity probe The glands are unscrewed by hand the probes inserted completely into the reactor until they rest on the...

Страница 12: ...Armfield Instruction Manual 8 CEM MKII Continuous Stirred Tank Reactor...

Страница 13: ...e top right hand corner of the screen as shown below Note that when operating the software for the first time it will be necessary to enable the USB virtual COM port by choosing the Red telephone icon...

Страница 14: ...e reactor The gland nut should be loosened prior to fitting the probe sensor then tightened to secure them CEXC fitted with CEM MkII Continuous Stirred Tank Reactor Connection to the Hot Water Circula...

Страница 15: ...EM software then choose the appropriate experiment from the menu For detailed information about operating the software refer to the section Operating the PC Software Display the mimic diagram by choos...

Страница 16: ...r of the unit should be turned on before hand Both the temperature controller and conductivity display should illuminate Filling the feed bottles Unscrew the feed bottle caps and pour appropriate reag...

Страница 17: ...and control of the heater stirrer and pumps Recorded results can be displayed in tabular and graph format The software runs on a WindowsTM PC which connects to the CEXC using a USB interface Installa...

Страница 18: ...e used with an optional Chiller CW 17 for operating temperatures below ambient The use of the heater or chiller is chosen in the initial menu when loading the CEM software When using the hot water cir...

Страница 19: ...ontinuously to correct for any error The hot water temperature T2 is chosen as the default process variable to provide a stable temperature T1 inside the reactor Control of T1 directly would be diffic...

Страница 20: ...ection for appropriate controller settings when using the CW 17 When using the Heater the temperature sensor T1 supplied with CEXC which is Reactor Temperature on the software should be set as the Pro...

Страница 21: ...erator a Indoor use b Altitude up to 2000m c Temperature 5 C to 40 C d Maximum relative humidity 80 for temperatures up to 31 C decreasing linearly to 50 relative humidity at 40 C e Mains supply volta...

Страница 22: ...is located on a printed circuit board inside the plinth on the right hand side This circuit is calibrated before despatch and should not require re calibration However should re calibration become ne...

Страница 23: ...ion Range 0 20 mS cm Fill a small beaker with a Conductivity standard solution e g 0 1M KCI giving a conductivity of 12 88mS at 25 C and measure the temperature of the standard solution using a suitab...

Страница 24: ...a reading of the Standard solution in the Low conductivity window on the software Remove the probe from the conductivity solution and rinse the probe in de mineralised water When the conditioning circ...

Страница 25: ...o sodium hydroxide conc in mixed feed mol dm3 a1 sodium hydroxide conc in reactor at time t mol dm3 sodium hydroxide conc in reactor after time mol dm3 b ethyl acetate conc mol dm3 same subscripts as...

Страница 26: ...er The conductivity of the reacting solution in the reactor changes with the degree of conversion and this provides a convenient method for monitoring the progress of the reaction either manually or b...

Страница 27: ...ate as follows Therefore add 9 79ml of concentrated Ethyl Acetate to 900ml of deionised or distilled water Shake the mixture vigorously until the two liquids have mixed Add further water to make up th...

Страница 28: ...ve the caps of the reagent bottles and carefully fill with the reagents to a level approximately 50mm from the top using a funnel if necessary Refit the caps Start the software using the option of the...

Страница 29: ...at the same molarity and temperature and a relationship has been established allowing conversion to be inferred from conductivity The calculations are best carried out using a spreadsheet such as EXCE...

Страница 30: ...calculate values of sodium hydroxide concentration a1 and sodium acetate concentration c1 and the degree of conversion Xa and Xc for each of the samples of conductivity taken over the period of the ex...

Страница 31: ...ate and the conversion Notes 1 It is recommended that this experiment should be repeated at various other temperatures to investigate the relationship between the specific rate constant k and the temp...

Страница 32: ...ree different temperatures is known is straightforward to apply the Arrhenius law and calculate the frequency factor and the activation energy values for Plotting Rate constant vs Temperature Obtentio...

Страница 33: ...e initial concentrations are equal both a0 and the conversion Xa then the concentrations are as follows NaOH CH3COOC2H5 C2H5OH CH3COONa a0 Xa a0 Xa Xa Xa Method Repeat Exercise A after removing the ba...

Страница 34: ...d this is achieved using the hot water circulator and setting the PID temperature controller in the software as detailed in previous experiments Start the water feed pump by setting the pump speed con...

Страница 35: ...e the Neperian logarithm of Plot vs time and calculate the slope straight line graph passing through the origin The slope is the average residence time tR which should be equal to where V is the react...

Страница 36: ...for a working volume of 1369ml and a working flow rate of 106 5ml min The difference between the theoretical and the experimental value may be caused by experimental errors such inaccuracies on the m...

Страница 37: ...West Street Ringwood Hampshire England BH24 1DY Tel 44 0 1425 478781 Fax 44 0 1425 470916 Email sales armfield co uk support armfield co uk Web http www armfield co uk US Office Armfield Inc 9 Trento...

Отзывы: