22
ON/OFF key again and wait. After the weight appears on screen, the message “Starting QuickConnect” appears on the top
part of the screen. Then you will see “Verifying Service”. After this message, you will see “Link Established. This means the
QuickConnect session was successful. If you see “Service Check Failed!” and “Device not found”, this means the
QuickConnect device is not in the vicinity, or in more advanced setting, the Local/Remote port is not right. If the session was
successful, the Bluetooth indicator will appear. Press PRINT key to send data to the device.
Manual
For a manual operation,
press ON/OFF and then press 5 (within 3 seconds)
. Immediately, the scale will scan the current
area to see if the device you defined in the Quick Connect screen is present. If it is present, it will connect to this device and
show the Bluetooth indicator. If the device is not found, then the appropriate status message will be shown. You may now
send and receive data from the device. Press PRINT key to send data to the device.
Quick Disconnect
If you want to disconnect from the device, you can either use the Quick Disconnect feature or go to the Bluetooth Setup
option and disconnect from there. To quick disconnect the device, use the same sequence that is described above; i.e.
press ON/OFF and then press 5 (within 3 seconds)
. This quick disconnection sequence can also be used if you have
connected to the device without going through Quick Connect.
Bluetooth Local Settings (ADVANCED)
This section introduces some of the advanced features available for the scale Bluetooth module. Most of these features
have to do with the local configuration of Bluetooth module. It is recommended that you are well versed with the above
described operation before you get into this section. Most of the time, you will not need to modify these settings.
Following are the local Bluetooth settings that can be modified to satisfy some specific requirements in your application. To
access these settings, in the Bluetooth Setup screen, scroll down “Bluetooth Local Settings.”
Bluetooth Enable
This menu option allows you to Enable or Disable the Bluetooth functionality. After toggling this option, the scale needs to be
restarted.
View Device Info
This menu option allows to view the Bluetooth modules name and device address.
Change Local Name
This menu option allows you to change the local name of the scale Bluetooth module. Currently, it is defaulted
to “ARLYN
SCALE”. You can modify this to anything you want depending on your needs.
Virtual Port Settings
This menu option specifies the characteristics of the aerial Virtual COM port through which to send out data. This setting, if
changed improperly, may disable your Bluetooth communication system.
Fixed Pin Settings
This shows the current Default Pin settings of the Bluetooth module. This allows the Bluetooth module to send this pin to any
device asking for a pin request. You will never need to enter the pin on the scale.
Entering this option, the default pin is shown on screen. The default pin is ‘0000’. You can press ENTER to change the
default pin. You have two options here:
1.
Default Fixed Pin:
You can put a default pin, of up to 16 alphanumeric characters in length. The next time a pairing is
needed, this pin has to be entered on the other remote device when a dialog box pops up.
2.
No Fixed Pin:
By pressing ENTER without entering a pin, the scale will assume that you want to enter the pin each
time a pairing is required. The next time you pair with a device, a dialog box will pop up on the scale screen for you
to enter a desired pin. That same pin has to be entered in the remote device as well.