background image

Chapter 3 Auxiliary Operations

On the menu screen, you can perform instrument settings.

 To access the menu

On the standby screen, hold down [<] for 3 seconds.

 Options on the menu

 Basic operations in the MENU screen

Change the MENU No. and set values.

[<] button: Moves the position of the selected digit (the digit that is blinking) 

back one space.

[   ] button: Finalizes the value that is being set (the value that is blinking), 

and move the position of the selected digit forward one space.

[>] button: Increases the value of the blinking digit by one. When the upper 

limit of the setting range is reached, the value returns to the 
beginning value.

 (Example: When the setting range is 0 to 9
                  0    1    2 • • • 8    9    0)

 Selecting a MENU item

Enter a MENU No. indicated in the above table to set that item or execute 
that function.

The following shows an example of how the date and time are set.

a

 On the standby screen, press and hold [<] for 3 seconds to display the 

MENU screen.
"1-0-0" will appear in the sub-screen with the "1" (digit) blinking.

b

 Press the [    ] button.

The "1" will stop blinking and the middle "0" digit will change to "1" and 
start blinking.

c

 Press the [>] button once.

The middle digit will change to “2”.

d

 Press the [    ] button.

The MENU No. will be finalized as "1-2" and the screen will transition to 
the time setting screen.

e

 Set the tens digit and the ones digit for the hour, and then set the tens digit 

and the ones digit for the minute, in that order.

 * Repeat the same steps as above to select and set the values of the other 

MENU items.

2.2.3 Measuring the Sample

a

 Firmly and securely attach the pipette tip to the end of the pipette.

b

 Aspirate the sample with the pipette.

c

 Press the [    ] button, and then dispense a drop of sample on the test strip.

Dispense the sample onto the sample dispense area of the test strip.

     Press the pipette button all the way in slowly.

d

 After you dispense a drop of sample, immediately press the [    ] button and 

insert the strip tray into the instrument. (Insert the strip tray within 10 
seconds of pressing the [    ] button in step 

d

.)

e

 After the measurement time elapses, the measurement results appear on 

the display.

f

 Check the measurement results.

g

 Withdraw the strip tray and take out the used test strip.

Check the test strip, and check that the sample has flowed in correctly and 
that there are no large air bubbles or foreign matter in the reaction cell. After 
checking, discard the test strip.

h

 The standby screen will appear 5 seconds after withdrawing the strip tray.

2.2.4 Measuring the control

a

 On the standby screen, press the [>] button. The control icon will appear.

b

 Dispense a drop of the control onto the test strip, and then perform the same 

procedure  as when measuring a sample.

c

 The measurement results are saved as a control measurement.

2.2.5 Understanding results

After measurement, the following information appears on the screen.

Upper line:The item name and measurement number appear alternately. 

If you press the     button, the display changes to the date and 
time.
If there is a status indication (*), the status appears instead of the 
measurement number.

Lower line:The measurement result value and units appear. However, if the 

status is UNDER or OVER, the measurement range appears.

(*) Status

UNDER  Under the lower limit of the measurement range.

Check the conditions. It is possible that drop dispensing failed and 
the sample did not enter the reaction cell, or a large bubble 
formed.

OVER 

Over the upper limit of the measurement range.

OVER* 

Over the upper limit of the measurement range, and the prozone 

effect or a reaction abnormality may have occurred.

HIGH 

The value is higher than the reference value. (only when a 

reference value is set)

LOW 

The value is lower than the reference value. (only when a 

reference value is set)

2.2.6 Checking the measurement result history

a

 On the standby screen, press [<]. 

The current date and time screen appears.

b

 Press [>]. 

The most recent test results appear.
Pressing [>] again displays the measurement results prior to the one 
currently displayed.

 MENU1

Option

Description

1. Date

Sets the date.

2. Time

Sets the time.

3. Date format

Sets the date format.
Selectable items: [YMD][MDY][DMY]

4. Sound volume

Sets the beeper volume.
Selectable items: 0 to 3

 MENU2

Option

Description

MENU No.

1 - 1
1 - 2

1 - 3

1 - 4

2 - ITEMx

2 - ITEMx
2 - ITEMx

2 - ITEMx

2 - ITEMx
2 - ITEMx

3 - 1

3 - 2 - 1

3 - 2 - 2

3 - 2 - 3

3 - 2 - 4

MENU No.

MENU No.

ITEM1 to
ITEM16

Reference value ON/OFF

Sets the reference value function to 
[ON] or [OFF].

2 - ITEMx

Unit

Sets the unit.
Selectable items: [ ][mmol/L][

μ

mol/L]

[mg/dL][g/dL][U/L]

High reference value

Sets the High reference value.

Low reference value

Sets the Low reference value.

Reagent in IC-CHIP 
Check parameter 
reference value

Parameter of reagent in IC-CHIP

User coefficient A
User coefficient B

 MENU3

Option

Description

1. Check measurement

Performs check measurement.
Checks light intensity of optical 
unit LED

Execute this when “E-130” appears.

Sets the threshold for scanning test 
strip barcodes

2. Optical
    unit

1. Check light intensity

2. Light absorbance 
    blank adjustment

3. Light intensity gain 
    adjustment

4. Barcode threshold 
    setting

Содержание Lab 004

Страница 1: ... the instrument and the proper procedures for operation and maintenance Follow the instructions in this operating manual in order not to defeat the purpose of the protective features of the instrument For the purchase of reagents consumables or other optional items contact your distributor TAKE THE UTMOST CARE WHEN HANDLING BLOOD This instrument uses blood as sample Blood may be contaminated with ...

Страница 2: ...Dual wavelength transmitted light measurement Number of test strips measurable at once 1 Measurement time Listed in package insert of reagent Approximately 4 μL Sample consumption Listed in package insert of reagent 1 sample Approximately 10 minutes in environments 25 C and below Memory capacity 50 measurements Display Segment LCD Flat keypad Micro USB type B 110 W 225 D 54 H mm Weight Approximate...

Страница 3: ... battery in place of the provided AC adapter Use a battery with a rating of 5 V and 2 0 2 4 A Do not disassemble the instrument Do not modify the instrument Disassembly and modification of the instrument may result in exposure to pathogenic microorganisms or cause fire or damage to the instrument and consequently lead to personal injury 1 4 2 Precautions for Instrument Relocation Before relocating...

Страница 4: ...ior to measurement Do not store a sample in direct sunlight or a location directly exposed to air flow from an air conditioner or other source Ensure the instrument is in a proper environment before turning on the power see 1 4 1 Precautions for Instrument Installation Do not place containers or bottles that contain liquid on the instrument Sample or other liquid that gets inside the instrument ma...

Страница 5: ...econds after withdrawing the strip tray 2 2 4 Measuring the control a On the standby screen press the button The control icon will appear b Dispense a drop of the control onto the test strip and then perform the same procedure as when measuring a sample c The measurement results are saved as a control measurement 2 2 5 Understanding results After measurement the following information appears on th...

Страница 6: ...ace the parts at recommended intervals 1 Clean the strip tray At the end of the day When measurement for the day is complete be sure to clean the strip tray Prepare cotton swabs distilled water soft cloth alcohol and disposable gloves Use a soft cloth or cotton swab moistened with water or alcohol to clean the strip tray 2 Cleaning the photometric window when E 130 occurs Prepare cleaning pad dist...

Страница 7: ...ur distributor 84 07200A Rev OCT 2019 T 999 A problem occurred in communication with an external device Connect the external device with the communication cable correctly E 102 The date and time settings are incorrect Set the date and time correctly E 103 The photometric window is dirty Clean the photometric window E 122 The ambient temperature of the instrument is less than 10 C or over 35 C Adju...

Отзывы: