1.4.1 Precautions for Instrument Installation

Before installation of the instrument, read the following notes and always take 
proper safety precautions.

• Determine a location for the instrument and install it in that location. Hold 

the bottom of the instrument with both hands when carrying it.

• Install the instrument on a surface with sufficient clearance around the 

instrument. Do not block the vents on the back of the instrument.

• Install the instrument where temperature and humidity can be maintained 

in the following ranges:

  Temperature: 10 - 35°C
  Humidity: 20 - 80%
  Installation in an environment outside these ranges may cause inaccurate 

measurement results.

• Install the instrument on a level, sturdy, vibration-free platform. Operation 

of the instrument in an unstable place may cause trouble with or 
malfunction of the instrument resulting in personal injury. Do not install the 
instrument where it may fall off or topple over.

• Do not use the instrument in a location where dust has accumulated. 

There is a risk of current leakage.

• Do not install the instrument near:
       - Places that store chemicals
       - Equipment that generates corrosive gas or electrical noise or

      - Equipment that may bring the instrument out of the operating  
        temperature and humidity ranges.

These factors may cause trouble with or malfunction of the instrument, 
resulting in personal injury, or may otherwise cause inaccurate 
measurement results.

• Install the instrument in a place where condensation, intense light such as 

direct sunlight or wind can be avoided. These factors may cause 
inaccurate measurement results, as well as deformation of or damage to 
the instrument.

• Apply the correct voltage (100 - 240 V AC ± 10%) and frequency (50/60 Hz) 

to the instrument. Improper voltage and frequency may result in fire or 
damage to the instrument and consequently lead to personal injury.

• Use the power cord and AC adapter that come with the instrument. 

• Connect the AC adapter directly to a single outlet. The power supply for 

the instrument is 33 VA. If you use a power strip, make sure it is 
grounded. Also, check if the power capacity for the power strip is within 
the specified range.

• It is possible to use a commercially available mobile battery in place of the 

provided AC adapter. Use a battery with a rating of 5 V and 2.0 - 2.4 A.

• Do not disassemble the instrument. Do not modify the instrument. 

Disassembly and modification of the instrument may result in exposure 
to pathogenic microorganisms or cause fire or damage to the instrument 
and consequently lead to personal injury.

1.4.2 Precautions for Instrument Relocation

Before relocating the instrument, read the following notes and always take 
proper safety precautions.

• Hold down the      button to turn off the power, and then disconnect the 

power cable from the instrument.

• Remove the test strip from inside the instrument. There is a risk of 

pathogenic microbes scattering inside the instrument if the instrument is 
moved with a test strip still inside.

• If the instrument will be moved a long distance, secure the strip tray and 

pipette guide with tape to ensure that neither opens or comes off.

1.4  Installation

• There is a risk of instrument failure if moved without securing the strip tray 

and pipette guide. Place the instrument in the dedicated box together with 
the accessories to move.

• Do not subject the instrument to shock or vibration when moving. There is 

a risk of instrument failure.

• Before moving the instrument, read “1.4.1 Precautions for Instrument 

Installation”.

1.4.3 Installing the Instrument

a

 Remove the fixing tape from the two positions.

b

 Connect the power cord and AC adapter.

c

 Insert the power cord into the power input terminal socket on the 

instrument.

d

 Plug the AC adapter into an outlet.

e

 Attach the pipette guide to the pipette guide attachment rails on the 

instrument.

NOTE:

The pipette guide can be attached to either the left side or the right side 
of the instrument.

1.4.4 Turning on the Power

a

 Make sure there is no test strip on the strip tray, then insert the strip   

tray into the instrument.

b

 Press and hold     for 1 second.

Warm-up will complete in about 10 minutes.

c

 Ensure that the standby screen is displayed.

If the standby screen does not appear and "RE-ADJ?" is displayed, press 
the [    ] button. Automatic adjustment of the optical unit will be performed. 
You can also press the [>] button to go to the standby screen without 
performing automatic adjustment. In this case, measurement cannot be 
performed.

1.4.5 Turning off the Power

a

 Ensure that the standby screen is displayed.

b

 Press and hold     for 3 seconds.

• The power will turn off.

Pipette guide

Содержание Lab 004

Страница 1: ... the instrument and the proper procedures for operation and maintenance Follow the instructions in this operating manual in order not to defeat the purpose of the protective features of the instrument For the purchase of reagents consumables or other optional items contact your distributor TAKE THE UTMOST CARE WHEN HANDLING BLOOD This instrument uses blood as sample Blood may be contaminated with ...

Страница 2: ...Dual wavelength transmitted light measurement Number of test strips measurable at once 1 Measurement time Listed in package insert of reagent Approximately 4 μL Sample consumption Listed in package insert of reagent 1 sample Approximately 10 minutes in environments 25 C and below Memory capacity 50 measurements Display Segment LCD Flat keypad Micro USB type B 110 W 225 D 54 H mm Weight Approximate...

Страница 3: ... battery in place of the provided AC adapter Use a battery with a rating of 5 V and 2 0 2 4 A Do not disassemble the instrument Do not modify the instrument Disassembly and modification of the instrument may result in exposure to pathogenic microorganisms or cause fire or damage to the instrument and consequently lead to personal injury 1 4 2 Precautions for Instrument Relocation Before relocating...

Страница 4: ...ior to measurement Do not store a sample in direct sunlight or a location directly exposed to air flow from an air conditioner or other source Ensure the instrument is in a proper environment before turning on the power see 1 4 1 Precautions for Instrument Installation Do not place containers or bottles that contain liquid on the instrument Sample or other liquid that gets inside the instrument ma...

Страница 5: ...econds after withdrawing the strip tray 2 2 4 Measuring the control a On the standby screen press the button The control icon will appear b Dispense a drop of the control onto the test strip and then perform the same procedure as when measuring a sample c The measurement results are saved as a control measurement 2 2 5 Understanding results After measurement the following information appears on th...

Страница 6: ...ace the parts at recommended intervals 1 Clean the strip tray At the end of the day When measurement for the day is complete be sure to clean the strip tray Prepare cotton swabs distilled water soft cloth alcohol and disposable gloves Use a soft cloth or cotton swab moistened with water or alcohol to clean the strip tray 2 Cleaning the photometric window when E 130 occurs Prepare cleaning pad dist...

Страница 7: ...ur distributor 84 07200A Rev OCT 2019 T 999 A problem occurred in communication with an external device Connect the external device with the communication cable correctly E 102 The date and time settings are incorrect Set the date and time correctly E 103 The photometric window is dirty Clean the photometric window E 122 The ambient temperature of the instrument is less than 10 C or over 35 C Adju...

Отзывы: