1.1.1 Features

■ Compact size enables installation in almost any location

This compact instrument is close to A5 in size, and weighs less than 1 kg for 
easy  carrying.  Water  supply  and  drainage  equipment  is  not  needed,  so  the 
instrument  can  be  easily  installed  in  any  location  and  used  for  immediate 
testing  during  physical  examinations  and  in  clinics.  This  instrument  can  be 
used  to  measure  multiple  biochemical  items  that  are  useful  for  monitoring 
internal disease and other conditions, such as creatinine.

■ Easy calibration by IC-CHIP

Calibration  can  be  performed  without  the  need  for  a  calibrator  using  the 
IC-CHIP provided with each test strip. When opening a reagent, simply read 
the IC-CHIP into the instrument. This enables automatic calibration, even for 
differences between lots.

■ Identification of reagent type by barcode

When  measurement  is  started  the  type  of  reagent  loaded  is  automatically 
identified.  Multiple  reagent  types  can  be  measured  without  the  need  for 
additional operation.

1.1 Overview

1.2  Unpacking

1.3  Part Names and Functions

Chapter 1 Before Use

1.3.1  Front View

1.3.2  Rear

Name

Configuration

Required sample 
volume

Number of measurement 
samples

Startup time

Operation buttons

External output

Dimensions

Altitude

Pollution degree

Measurement 
environment
Storage 
environment

Environment during 
transport

The Lab 004

Instrument, accessories

Measurement objects

Plasma or serum

Supported test strips

The Lab 004 test strips

Measurement items

Measurement ranges

Listed in package insert of reagent

Measurement principle

Listed in package insert of reagent

Dual-wavelength transmitted light measurement

Number of test strips 
measurable at once

1

Measurement time

Listed in package insert of reagent

Approximately 4 μL

Sample consumption

Listed in package insert of reagent

1 sample

Approximately 10 minutes 
(in environments 25°C and below)

Memory capacity

50 measurements

Display

Segment LCD

Flat keypad

Micro USB type B 

110 (W) 

×

 225 (D) 

×

 54 (H) mm

Weight

Approximately 1 kg

Power requirements

100 - 240 V AC, 50/60 Hz (with AC adapter)

Location of use

For indoor use only

Up to 2000 m

2

Power consumption

AC adapter

Maximum 33 VA

5 V DC

Sound pressure level

Less than 80 dB

Temperature: 10 - 35°C
Humidity: 20 - 80% RH (No condensation)
Temperature: 1 - 35°C
Humidity: 20 - 80% RH (No condensation)

Temperature: -10 - 60°C
Humidity: 20 - 80% RH (No condensation)

Over voltage category

Expected life

II

5 years (According to company data)

Temporary over voltage Short-term: 1440 V

Long-term: 490 V

1.1.2 Specifications

1.1.3  Measurement Principle

■ Test strip measurement

The instrument is equipped with multiple light source LEDs of varying 
wavelength. The light is dispersed on the light path and transmitted to the 
photometric unit. The dispensed sample fills the photometric cell and reacts 
with the reagents in the cell, changing the absorbance and turbidity. The light 
source LED shines on the cell, and the light that transmits through is read by 
a photodiode (receiver). The amount of transmitted light that is read is 
converted to a concentration.

Unpack the box and make sure you have all items listed in this section. If 
anything is missing or damaged, contact your distributor.
• Instrument
• Power cord
• AC adapter
• Operating manual
• Pipette

c

 Pipette guide

    attachment rails

a

 Display

b

 Operation 

     buttons

d

 Strip tray

e

 IC-CHIP 

     insertion port

f

 USB port

g

 Power input 

     terminal

Содержание Lab 004

Страница 1: ... the instrument and the proper procedures for operation and maintenance Follow the instructions in this operating manual in order not to defeat the purpose of the protective features of the instrument For the purchase of reagents consumables or other optional items contact your distributor TAKE THE UTMOST CARE WHEN HANDLING BLOOD This instrument uses blood as sample Blood may be contaminated with ...

Страница 2: ...Dual wavelength transmitted light measurement Number of test strips measurable at once 1 Measurement time Listed in package insert of reagent Approximately 4 μL Sample consumption Listed in package insert of reagent 1 sample Approximately 10 minutes in environments 25 C and below Memory capacity 50 measurements Display Segment LCD Flat keypad Micro USB type B 110 W 225 D 54 H mm Weight Approximate...

Страница 3: ... battery in place of the provided AC adapter Use a battery with a rating of 5 V and 2 0 2 4 A Do not disassemble the instrument Do not modify the instrument Disassembly and modification of the instrument may result in exposure to pathogenic microorganisms or cause fire or damage to the instrument and consequently lead to personal injury 1 4 2 Precautions for Instrument Relocation Before relocating...

Страница 4: ...ior to measurement Do not store a sample in direct sunlight or a location directly exposed to air flow from an air conditioner or other source Ensure the instrument is in a proper environment before turning on the power see 1 4 1 Precautions for Instrument Installation Do not place containers or bottles that contain liquid on the instrument Sample or other liquid that gets inside the instrument ma...

Страница 5: ...econds after withdrawing the strip tray 2 2 4 Measuring the control a On the standby screen press the button The control icon will appear b Dispense a drop of the control onto the test strip and then perform the same procedure as when measuring a sample c The measurement results are saved as a control measurement 2 2 5 Understanding results After measurement the following information appears on th...

Страница 6: ...ace the parts at recommended intervals 1 Clean the strip tray At the end of the day When measurement for the day is complete be sure to clean the strip tray Prepare cotton swabs distilled water soft cloth alcohol and disposable gloves Use a soft cloth or cotton swab moistened with water or alcohol to clean the strip tray 2 Cleaning the photometric window when E 130 occurs Prepare cleaning pad dist...

Страница 7: ...ur distributor 84 07200A Rev OCT 2019 T 999 A problem occurred in communication with an external device Connect the external device with the communication cable correctly E 102 The date and time settings are incorrect Set the date and time correctly E 103 The photometric window is dirty Clean the photometric window E 122 The ambient temperature of the instrument is less than 10 C or over 35 C Adju...

Отзывы: