The third module presented on Waller’s home screen is the temperature
module. There we can see the following options:
Sensor. Temperature.
Notes: If the user wants to assemble a cable with other standards or
perform the impedance conference, follow these steps:
1. Put the multimeter on the resistance scale
2. Connect the black lead to the ground pin of the P10 connector
3. Connect the red lead to the + Signal, to measure YSI 400 and YSI 700 A.
4. Connect the red lead to - Signal to measure YSI 700 B.
Check the values according to the previous table.
Settings
Sensor:
option to select between the 400 series or 700 series standard.
Temperature:
option to select the temperature.