18
It is necessary to assemble the test circuit to carry out the leak test, as
follows:
1. Connect a T (adapter) between the pressure line to be tested;
2. On one side of the T (adapter), connect a hand pump system;
3. On the other, connect the device or system to validate leakage;
4. Connect the third outlet to Waller.
Leakage Test Assembly