Electromagnetic Compatibility
0
0
1
.20
05
8.E
N
re
v.
10
Electromagnetic Compliance
The SARA LITE has been tested for compliance with current regulatory standards regarding its capacity
to block EMI (electromagnetic interference) from external sources.
Nonetheless, some procedures can help reduce electromagnetic interferences:
•
Use only ArjoHuntleigh cables and spare parts to avoid increased emissions or decreased immunity
which can compromise the correct functioning of the equipment.
•
Ensure that other devices in patient-monitoring and/or life-support areas comply to accepted emissions
standards.
•
Maximize the distance between electro-medical devices. High-powered devices may produce EMI that
can affect the lift. Refer to separation distance table further on in this document.
For more information on how to manage the unit’s RF electromagnetic environment, please consult the
AMI TIR 18-1997 - Guidance on Electromagnetic Compatibility of Medical Devices for Clinical/Biomedical
Engineers
.
Electromagnetic Emissions
Guidance and Manufacturer’s Declaration -
Electromagnetic Emissions - For all Equipment and Systems
The SARA LITE is intended for use in the electromagnetic environment indicated below. The customer
or the user of the SARA LITE should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The SARA LITE uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The SARA LITE is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Not applicable
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Not applicable