10 AH-MIN V3
08
/
10
/17
General
Examine the bed for obvious signs of damage. Make sure that all nuts, bolts, clamp screws and
other fastenings are tight and not missing.
Examine the flexible cables for cuts, abrasions, kinking or other deterioration. Ensure that the
power supply connector plug is securely attached and not damaged. If appropriate, check that the
plug fuse is of the correct type (BS 1362 5amp)
If either the plug or cable is damaged then both the plug and cable must be replaced as a complete
assembly by an approved service agent.
6. Cleaning
Warning
Before cleaning, disconnect the bed from the mains power supply.
Gloves and protective clothing should always be worn
when carrying out cleaning procedures.
All surfaces can be cleaned by wiping with a soft cloth moistened with hot water and a mild
detergent. Avoid splashing the actuators and handsets.
Rinse with a cloth moistened with clean water and dry with paper towels.
To remove potentially infectious material such as body fluids, or when used by a client with a known
infection, clean afterwards with NaDCC (ie Presept, Actichlor) at 10,000 ppm of
available chlorine.
Rinse and dry.
Allow cleaned parts to dry thoroughly before replacing the mattress.
Any proprietary furniture polish may be used on the head and foot panels.
Caution
Do not use abrasive compounds or pads.
Do not use hypocarbonate or phenolic cleaning solutions
7. Maintenance / Troubleshooting
Warning
Disconnect the bed from mains power supply before starting any maintenance activity.
The following preventative maintenance procedures should be carried out at intervals of 12 months
or as specified.
All service and repair activities must only be performed by properly qualified and trained technicians
approved by ArjoHuntleigh.
Содержание Minuet 100500AUS