Minuet® 

USER MANUAL 

 

Содержание Minuet 100500AUS

Страница 1: ...1 AH MIN V3 08 10 17 Minuet USER MANUAL ...

Страница 2: ...uld result in death injury or other serious adverse reactions Cautions given in this manual identify possible hazards in procedures or conditions which if not correctly followed could result in equipment damage or failure Notes given in this manual are used to explain or amplify a procedure or condition WARNING The tilt facility if available should only be used with the guidance of a clinically qu...

Страница 3: ... use and maintenance of the ArjoHuntleigh Minuet bed with model numbers listed in the table below 1 Introduction 2 Clinical Applications Area of Use Private nursing and convalescent homes Indications The use of the bed is to assist in patient positioning whilst reducing lifting and handling stresses for the carer The four section deck allows the patient to be moved to an upright sitting position w...

Страница 4: ...right when assembled Use carpet or other suitable protection to avoid damage 1 On both the head and foot end column assemblies turn the castors upward and apply the brakes refer to Operations section Lay the foot end flat on the floor with the spigots on the frame upper most WARNING Heavy equipment hazard The bed and its subassemblies are heavy and appropriate precautions should be taken to avoid ...

Страница 5: ...tand the assembled components on one side with the connector sockets on the backrest actuator or control box A uppermost 4 Assemble the head end deck assembly to the foot end deck assembly 5 Assemble the foot end to the deck A ...

Страница 6: ...tuator cable through the support loop C on the frame and connect it to socket 3 D on the backrest actuator control box Adjust the cable coils in the loop so the cable is under tension but not strained Note Backrest sheet metal top must be removed by unclipping two tabs E in order to access the control box cable ports 8 On models with electric leg elevation carefully thread the leg actuator cable u...

Страница 7: ...e rails to lift or move bed WARNING Make sure that the mains supply cable is not stretched when connected Make sure that the handset and mains cable cannot become entangled with moving parts of the bed Lower the bed to minimum height and disconnect the bed from mains power supply On both the head and foot end column assemblies turn the castors inward toward each other and apply the brakes Turn the...

Страница 8: ...d is stationary When operating the bed in tilt mode apply the brakes at one end only Do not expose the equipment especially the mattress to naked flames such as cigarettes etc Use only ArjoHuntleigh accessories designed or approved for use with the bed o When moving bed wrap electrical power cable around Cable Tidy located at bed head end in order to avoid damage to cable o Ensure the bed is conne...

Страница 9: ...ntrol handset the legs are elevated by increasing the angle of the thigh section When the legs are raised the Kneebreak angle can be manually adjusted by raising or lowering the foot end which is supported by a self locking ratchet strut To adjust hold the mattress retention rail and lift the foot end slightly to disengage the latch then raise or lower it reversing the movement a little to re enga...

Страница 10: ...ened with hot water and a mild detergent Avoid splashing the actuators and handsets Rinse with a cloth moistened with clean water and dry with paper towels To remove potentially infectious material such as body fluids or when used by a client with a known infection clean afterwards with NaDCC ie Presept Actichlor at 10 000 ppm of available chlorine Rinse and dry Allow cleaned parts to dry thorough...

Страница 11: ...n one 1 minute to complete 1 Visual inspection of handset casing 1 1 Inspect the outer plastic casing for any obvious cracks or damage 1 2 Inspect for damage to the front cover outer label particularly over the buttons where the label may be torn or indented If any damage is identified while conducting checks 1 1 and 1 2 immediately quarantine the bed and report the damage to maintenance staff or ...

Страница 12: ...ace the Handset otherwise contact an ArjoHuntleigh approved Technician The following table identifies some fault symptoms their possible cause and suggest some simple checks and corrective actions If these actions fail to solve the problem consult the ArjoHuntleigh Service Department on Freecall 1800 072 040 Symptom Possible Cause Action Actuators fail to work Power disconnected or batteries disch...

Страница 13: ...s from the date of purchase against fault arising due to defects in manufacture or materials Parts replaced or repaired under the terms of this warranty will be covered for the balance of the 36 month period The warranty applies only to parts made or approved by ArjoHuntleigh The warranty does not extend to items or components which may need replacement due to normal wear and tear eg Castors prote...

Страница 14: ...ndard 20kg Head End Assembly standard 17kg Foot End Deck Assembly with leg elevation 17kg Head End Deck Assembly adjustable Backrest 21kg Maximum Safe Working Load incl Accessories 180kg Electrical Data Power Input 1A max at 230V a c 50 60Hz Power Output 24V DC Duty Rating Intermittent 10 Max 2min 18min Electrical Safety Standard Complies with EN 60601 1 UL60601 1 Electrical Shock Protection Class...

Страница 15: ...Ext BL Cap 19Dia Footend Base Assembly 2 FPL2655509 Bumpons SJ 5508 Headend Base Assembly 2 FPL5464023 22mm Sq Int BL Caps B rest Th panel F rest panel 4 WSH11000092 10mm White Nylon Washes B rest Th panel F rest panel 4 FMT2218010 10mm Axle Caps B rest Th panel F rest panel 4 FMT9991703 Rastomat Cathces Footrest Panel Assembly 2 MCC2699093 White Nylon Bushes Column Base Assembly 2 MCC2699100 Whit...

Страница 16: ...ISBANE As our policy is one of continuous improvement we reserve the right to modify designs without prior notice ArjoHuntleigh ABN 75 007 200 900 A member of the Huntleigh Technology PLC Group of Companies Acare Medical Science Co Ltd No 6 Hongxingyi Road Hongwan Industrial Park Zhuhai Guangdong P R China 519060 TEL 86 756 6326076 FAX 86 756 6326103 ...