Using your MAXI MOVE
17
001
.25
06
0.E
N
re
v.
1
1
Testing the Attachment
To ensure that the attachment is safely connected
and secured to the T-bar, hold the attachment
firmly with both hands, without pushing on the
locking clip thumb pads, and lift the attachment
upwards firmly (see Fig. 16). If the attachment
becomes dislodged from the T-bar, DO NOT use
the MAXI MOVE. Contact your local
ArjoHuntleigh agent.
Fig. 16
WARNING: DO NOT lower the attachment onto
rigid surfaces (e.g. bed, floor, wheelchair
armrests, etc.), to avoid the possibility of the
attachment becoming dislodged from the T-bar.
A dislodged attachment may later detach
completely from the unit, causing the patient to
fall.
WARNING: Always check that the attachment is
locked in place by: 1) verifying the markings on
the attachment that show that there is a proper
alignment, and 2) verifying that the locking clip is
fully engaged, thus ensuring that the attachment
is securely fastened.
Fig. 15
Markings on
the attachment
are aligned.
Locking clip is
fully engaged;
is fastened
securely.
CORRECT
Markings on
the attachment
are misaligned.
INCORRECT
Locking clip is
NOT engaged;
the attachment
is not fastened
securely.
CORRECT
INCORRECT
the attachment
WARNING: Always check that the attachment is locked in place
by: 1) verifying the markings on the attachment that show that
there is a proper alignment, and 2) verifying that the locking clip
is fully engaged, thus ensuring that the attachment is securely
fastened. An unsecured spreader bar may lead to a patient fall.