Service Procedures
MRM Malibu/Sovereign
26
SP 15 Perform a Full Feature
Functionality Test
Final service check
Function
Activation
Demands for approval
Shower handle and holder
No leaks or cracks
All functions
Buttons on tub and
hand control
Activation/deactivation
Hi/lo full stroke
No noise or "jumps"
Temp setting shower max
Temp setting fill max
Mechanical scalding
protection
41°C -1°C (106°F-1.8°F)
44°C -1°C
(111°F-1.8°F)
Autodisinfection
Button on tub panel
to be pressed for 5
seconds
Chair raising from tub bottom to 300 mm.
(15sec.) if not already up
Disinfection air-lines (10sec.)
Reaction time in pump :120 sec. when shor-
tend by pressing button twice else 155sec.
Rinse (40sec.)
Blower running (40sec.)
Disinfection handle
Button on tub panel
to be pressed for 5
seconds plus trigger
pulled on handle
With autodisinfection:
Flow can be shut off by pressing the button
(Flow -min 120 sec. when interrupted on
button,-max 155 sec. without interruption.
With manual disinfection only:
Flow can be shut off by pressing the button
(Flow: time-out 5 minutes.)
LED indication on tub con-
trol panel
Connected to mains
A:
Steady green or flashing green
Connected to mains
B:
steady green
Lock button active
C:
steady yellow
Disinfection button
activated.
D:
steady green (level OK)
D:
flashing green and beeping (level low)
Disinfection button
deactivated
D:
flashing green (with autodisinfection)
D:
No light (with manual disinfection)
Scalding protection
active: Water temp.
>45° C for 10 sec.
E:
flashing yellow
Lift out activated on
hand control but tub
is not in lowest posi-
tion
(only with Hi/lo)
F:
flashing yellow
In Service mode:
DIP switch 3 in pos
OFF
DIP switch 4 in pos
OFF
E:
flashing yellow
Disconnected from
mains
A:
Battery status
Steady green= Battery OK
A
B
C
D
E
F