20
11. Technical Specification
PUMP
Model:
Flowtron Universal
Part Numbers:
507001
507009AU
UK
Australia
Pressure Range:
Calf/Thigh: 40 mmHg ± 5 mmHg
Foot: 130 mmHg ± 10 mmHg
Supply Voltage:
100-230 V
Supply Frequency:
50-60Hz
Power Input:
38-77 VA
Size:
230 x 270 x 150 mm
Weight:
3.9 kg
Case Material:
Fire Retardant ABS Plastic
Plug Fuse Rating:
5A to BS1362 (UK ONLY)
Degree of protection
against electric shock:
•
Mains Connected: Class II, Double Insulated with Functional Earth
Type BF
•
Mains Not Connected: Internally Powered
Degree of protection
against liquid ingress:
IPX0 - No protection
Mode of operation:
Continuous
PUMP SYMBOLS
The operator must read
this document
(Instructions for Use)
before use.
Note: This symbol is blue
on the product label.
Run/Standby.
Note: Unit is not
isolated from
mains supply.
Do not dispose of in
domestic refuse
With respect to electric
shock, fire and
mechanical hazards only
in accordance with CAN/
CSA-C22.2 No. 60601.1
(2008).
MEDICAL EQUIPMENT
Double Insulated
Type BF
Refer to this document
(Instructions for Use) for
a description of the
product classification (3rd
Edition).
SN:
Serial Number
Ref:
Model number
Refer to this document
(Instructions for Use) for
a description of the
product classification
(2nd Edition).
Manufacturer: This
symbol is
accompanied by
the name and the
address of the
manufacturer.
i
Содержание FLOWTRON UNIVERSAL
Страница 1: ...FLOWTRON UNIVERSAL Instructions for Use with people in mind 0086 ...
Страница 2: ......
Страница 4: ... ii ...
Страница 30: ...24 ...