If the pump is stored in conditions outside of the “Operating” ranges, it should be allowed time to stabilise at
normal operating conditions before use.
ENVIRONMENTAL INFORMATION
Condition
Temperature Range
Relative Humidity
Atmospheric Pressure
Operating
+10 °C to +40 °C
(+50 °F to +104 °F)
30% to 75%
(non-condensing)
700 hPa to 1060 hPa
Storage and Transport
(Long Term)
+10 °C to +40 °C
(+50 °F to +104 °F)
20% to 95%
(non-condensing)
700 hPa to 1060 hPa
Storage and Transport
(Short Term)
-20 °C to +50 °C
(-4 °F to +122 °F)
20% to 95%
500 hPa to 1060 hPa
Guidance and manufacturer’s declaration - electromagnetic emissions
The pump is intended for use in the electromagnetic environment specified below. The customer or the
user of the pump should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR - 11
Group 1
The pump uses RF energy only for its internal
function. therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR - 11
Class A
The pump is suitable for use in all establishments
other than domestic and those directly connected
to the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies