11
After a few seconds, the following screen is
displayed, showing “Battery OK”:
If the text at the top of the screen says “Low
Battery” then you must connect mains power before
you can start therapy.
• At the end of the diagnostic test, the following screen
is displayed, showing the “Hours Run” at the top and
the software version of the pump at the bottom:
The “Hours Run” value displayed on this screen is
for service use only, and is different to the “Patient
Hours Meter” displayed during therapy. Refer to
“Controls, Alarms and Indicators” on page 6 for
the “Patient Hours Meter”.
Standby Screens
The “No Garments” Standby screen is then displayed:
There are two “leg” icons with no garments attached and
the text “Attach garments” at the top of the screen; this
indicates that no garments are currently connected to
either tubeset.
There is no text above the Start/Stop button until at
least one garment is connected to a tubeset.
Apply the prescribed ArjoHuntleigh garment(s) to the
patient by following the instructions included in the
garment packaging.
Battery OK
XX.XXXX-XXX
Hours Run: 145
Attach garments
Options
“Mains power
connected” icon
Battery
Indication
Содержание Flowtron ACS800