14
2. Foot garment 1 is inflated to 130 mmHg, with a
ramp up and hold time of 3 seconds. The foot
garment indicator is black while it is inflated:
3. The foot garment is deflated to zero. Both garment
indicators are grey:
4. Foot garment 2 is then inflated to 130 mmHg, with
a ramp up and hold time of 3 seconds. The foot
garment indicator is black while it is inflated:
5. The foot garment is deflated to zero. Both garment
indicators are grey:
6. This cycle of alternate garment inflations repeats
from step 2 (above) until the therapy is stopped.
If one or two calf (or calf & thigh) garments are attached
to the pump, then each garment is inflated to 40 mmHg,
with a ramp up and hold time of 12 seconds.
If a foot garment and a calf (or calf & thigh) garment are
attached to the pump, then since the inflate and hold
Stop
2 h
Stop
Stop
Stop