background image

001-04100-EN rev

. 2 

Troubleshooting

6

Technical Specifications

Troubleshooting

Care and Maintenance

PROBLEM

TO CHECK

The Turntable does not move 

when the cord is pulled.

•  Verify that the cord is correctly installed into the 

device.

•  Try to pull the other cord.

There is a rough transition between 

the fixed track and the Turntable.

•  Verify the gap between the tracks (1 mm to 1.6 mm).

•  Contact your local representative.

Preventive Maintenance Schedule

The 

BHM Turntable

 is subject to wear and tear, and the following actions must be performed when specified 

to ensure that the product remains within its original manufacturing specification.

WARNING

: The points on this checklist are the minimum the manufacturer recommends. In some 

cases more frequent inspections should be carried out. Continuing to use the 

BHM Turntable

 without 

conducting  regular  inspections  will  seriously  compromise  the  user  and  resident  /  patient’s  safety. 

Preventive maintenance specified in this manual can prevent accidents.

WARNING: 

Safety  related  maintenance  and  authorized  service  must  be  carried  out  by  qualified 

personnel, fully trained in servicing procedures by ArjoHuntleigh, and equipped with correct tools and 

proper documentation, including Parts List and Service Manual. Failure to meet these requirements 

could result in personal injuries and / or unsafe equipment.

POINTS TO BE INSPECTED BY  

USER / SERVICE TECHNICIAN 

FREQUENCY

Annually or 1500 cycles

Before every use

Initially  

  

Inspect for evidence of external damage or missing parts.

Make sure the lift moves smoothly between the 

BHM Turntable

 pivoting track and fixed tracks (tracks 

are well aligned and space between them is inferior to 1.6 mm [1/16 in]).

Make sure end stoppers of the 

BHM Turntable

 are in place and tightened.

Inspect pull cords for wear.

Annual Inspection by an Authorized Technician

The 

BHM Turntable

 must be inspected by a qualified ArjoHuntleigh, technician on an annual basis 

to ensure the proper functioning of its components, as well as the safety of the patients and care 

givers that use the 

BHM Turntable

. Please contact your local representative to make arrangements.

NOTE

: If the product does not work as intended, immediately contact your local ArjoHuntleigh, 

representative for support.

Cleaning

To clean the 

BHM Turntable

, wipe it down with a damp cloth using warm water and mild detergent. 

CAUTION

: Do not use phenol, chlorine or any other type of solvent that may damage the 

finish.

To ensure a better rolling surface for the trolley wheels, clean the inside of the rail every four 

months. To do so, insert a damp cloth in the opening and slide it from one end of the rail to the 

other.

Содержание 700.042 Series

Страница 1: ...BHM Turntable Serie 700 042XX Instruction for Use 001 04100 EN rev 2 March 2014...

Страница 2: ...oHuntleigh group of companies ArjoHuntleigh 2014 As our policy is one of continuous improvement we reserve the right to modify designs without prior notice The content of this publication may not be c...

Страница 3: ...2 How to Use this Manual 2 Safe Working Load 2 Important Safety Directions 3 Shock Prevention 3 Environmental Advice 3 Homecare Environment Considerations 3 Symbols Used 3 General Symbols 3 Product D...

Страница 4: ...001 04100 EN rev 2 Table of Contents iv...

Страница 5: ...vice or repair the BHMTurntable If any unit is defective contact your local ArjoHuntleigh vendor Always ensure that the BHMTurntable is installed and put into service by an authorized ArjoHuntleigh co...

Страница 6: ...ife span has been exceeded Wear may cause the breakage of a part and lead to a patient fall The life span is the safe operating time period before a complete overhaul is required Following are factors...

Страница 7: ...tion herein Always ensure that The unit is used by trained staff All controls and safety features are used only as per the rules specified in this manual Never attempt to force a control on the BHMTur...

Страница 8: ...table Max capacity 272kg 600lb Indoor use only Made in Canada 001 02294 REV 2 2013 05 700 04200 ArjoHuntleigh AB Verkstadsv gen 5 241 38 Esl v SWEDEN XXXXXXXXX 7 SAMPLE Legend 1 Activation Cord 2 Acti...

Страница 9: ...ng the cords 4 Move the lift out of the turntable along the new path 5 In case of a malfunction pull down the activation arm to unlock the pivoting track and rotate the track manually 6 WARNING Take c...

Страница 10: ...d Service Manual Failure to meet these requirements could result in personal injuries and or unsafe equipment POINTS TO BE INSPECTED BY USER SERVICE TECHNICIAN FREQUENCY Annually or 1500 cycles Before...

Страница 11: ...NTAL CONDITIONS Ambient temperature range Operation 10 C to 40 C 14 F to 122 F Storage 40 C to 70 C Relative humidity range Operation 30 to 75 non condensing Storage 10 to 100 non condensing Atmospher...

Страница 12: ...001 04100 EN rev 2 Product Dimensions 8 Product Dimensions Dimensions Fig 3 460 mm 18 1 8 320 mm 12 5 8 100 5 mm 3 15 16 502 5 mm 19 3 4 SQ...

Страница 13: ......

Страница 14: ......

Страница 15: ...UNSTABLE LU5 5XF Tel 44 0 1582 745 800 Fax 44 0 1582 745 866 E mail international ArjoHuntleigh com ITALIA ArjoHuntleigh S p A Via di Tor Vergata 432 IT 00133 ROMA Tel 39 0 6 87426211 Fax 39 0 6 87426...

Страница 16: ...oHuntleigh GETINGE and MAQUET ArjoHuntleigh focuses on patient mobility and wound management solutions GETINGE provides solutions for infection control within healthcare and contamination prevention w...

Отзывы: