21
5.3.2 Service
table
Note!
The service table is only shown for information purposes.
WARNING!
Maintenance may only be performed by authorized
service technicians.
WARNING!
The machine is connected to the electricity supply
and some components are live.
The service table below shows recommended maintenance
intervals. In addition to system testing, we recommend that
appropriate cleaning tests and temperature validation are
performed according to EN ISO 15883.
Measure
Each year/
10,000 cycles
Every other
year/20,000
cycles
Time*
(minutes)
1
General
1.1 Check the cabling and connection points.
•
10
1.2 Check the function of the cooling fan. Clean as
required.
•
5
1.3
&KHFNWKDWWKHSDQHOODEHOLV¿UPO\Dႈ[HGLV
undamaged and legible.
•
2
1.4 Check that the door lock and breaker are
functioning properly.
•
5
1.5 Check that a goods placement sign and quick
help instructions have been put up.
•
1
1.6 Check error codes and the number of
processes.
•
2
2
The chamber
2.1
&KHFNWKDWWKHURWDWLQJQR]]OHVDUHPRYLQJ
freely. Clean as required.
•
20
2.2
&KHFNWKDWWKH¿[HGQR]]OHVDUHQRWFORJJHG
with grime and deposits. Clean as required.
•
20
2.3
&KHFNIRUOHDNDJHLQWKHQR]]OHDWWDFKPHQWV
and the hose connections to the chamber.
•
-
2.4 Check for proper operation of the temperature
sensor.
•
10
2.5 Check that the door seal and chamber seal are
tight. Clean or replace as necessary.
•
10
2.6 Check the mounting for the goods holder.
•
2