11
To Adjust the Pressure
Control Knob Position
The pressure control knob (2) is locked in position to prevent
accidental rotation.
To adjust the position of the pressure control knob:
1.
Lift the lock pin (A) to release the control knob.
2.
Rotate the control knob (B) whilst the lock pin is raised.
3.
Release the lock pin when the pressure control knob is in the
desired position to lock the control knob.
Initial Pressure Setting
4.
If you are using the
Alpha Trancell
Deluxe seat cushion, set
the pressure control to the maximum position (8).
5.
If you are using a mattress, adjust the pressure for the
appropriate patient weight. Refer to the patient weight guide
on the back of the pump for the correct setting. In the case of
a patient sitting in bed, increase the pressure control knob by
two settings.
Lock
Pin
A
B
NOTE
Rotate the pressure control knob clockwise to increase and
counterclockwise to decrease pressure.
WARNING
The weight guide is only an indication and should not replace clinical judgement.
kg
lbs
Alpha PREVENT
Trancell
II
Alpha Trancell
Deluxe
If Patient
Sitting in Bed
40
88
2
1
1
60
132
2-4
1-2
2-3
80
176
4-5
2-3
4
100
220
Trancell II
3+
5-6
+2