4

Intended Use

Repositioning Sling / Disposable Repositioning Sling

 

are products intended for assisted lateral repositioning 

and/or lateral transfer of patients/residents with limited 

ability to move.

 

The 

Disposable Repositioning Sling

 is intended to be 

used for a limited period only, and, by nature of its 

design, must be treated as a disposable and patient/

resident specific product.

The 

Repositioning Sling / Disposable Repositioning 

Sling

 shall only be used by appropriately trained 

caregivers with adequate knowledge of the care 

environment, and in accordance with the instructions 

outlined in the 

Instructions for Use (IFU)

.

 

The 

Repositioning Sling / Disposable Repositioning Sling

 

are intended to be used in hospital environments, nursing 

homes, other healthcare facilities and home care. 

The 

Repositioning Sling / Disposable Repositioning 

Sling

 should only be used for the purpose specified in 

this 

IFU

. Any other use is prohibited.

Patient/Resident Assessment

We recommend that facilities establish regular assessment 

routines. Caregivers should assess each patient/resident 

according to the following criteria prior to use:

• 

The patient/resident shall be passive,

•  Might be almost completely bed ridden,

• 

Often stiff, contracted joints,

• 

Totally dependent on the caregiver,

•  Physically demanding for caregiver,

•  Stimulation and activation is not a primary goal

Safe Working Load (SWL) for

:

•  MAA6000: 272 kg (600lbs) (

Repositioning Sling)

•  AHD001: 272 kg (600lbs) (

Disposable 

Repositioning Sling)

The right type and size of slings should be used after 

proper assessment of each patient/resident’s size, 

condition and the type of lifting situation

If the patient/resident does not meet these criteria an 

alternative equipment/system shall be used.

Expected Service Life

The expected service life of 

Repositioning Sling / 

Disposable Repositioning Sling

 is the maximum period 

of useful life.

The expected life of the sling is dependent on the 

actual use conditions. Therefore, before use, always 

make sure that the sling does not show any sign of 

fraying, tearing or other damage and that there is no 

damage (i.e. cracking, bending, breaking). If any such 

damage is observed, do not use the sling. If you have 

any doubts about sling safety, as a precaution and to 

ensure safety, do not use the sling. 

The expected service life:

•  Service Life: 2 years (

Repositioning Sling)

•  Service Life: 2 weeks (

Disposable Repositioning 

Sling)

•  Shelf Life: 5 years (

Repositioning Sling / 

Disposable Repositioning Sling)

Содержание AHD001

Страница 1: ...INSTRUCTIONS FOR USE Repositioning Slings 04 SQ 00 EN_7 11 2021...

Страница 2: ...the Instructions for Use Design Policy and Copyright and are trademarks belonging to the Arjo group of companies Arjo 2021 As our policy is one of continuous improvement we reserve the right to modify...

Страница 3: ...sident Assessment 4 Expected Service Life 4 Safety Instructions 5 Serious incident 5 Parts Designation 5 Preparations 6 Before First Use 6 steps 6 Before Every Use 7 steps 6 After every use 6 Sling Se...

Страница 4: ...e patient resident shall be passive Might be almost completely bed ridden Often stiff contracted joints Totally dependent on the caregiver Physically demanding for caregiver Stimulation and activation...

Страница 5: ...s being used WARNING To avoid injury never let the patient smoke when using the sling The sling can catch on fire WARNING To avoid injury only store the equipment for a short period of time If stored...

Страница 6: ...ling Check for fraying loose stitching thread breakage tears fabric holes soiled fabric damaged loops unreadable or damaged label WARNING To prevent cross contamination always follow the disinfection...

Страница 7: ...lbs One Size One Size One Size AHD001 272 kg 600 lbs One Size One Size One Size Lift Maxi Move SWL 125 kg 275 lbs 227 kg 500 lbs Spreader Bar 2 point loop Small 2 point loop Medium 2 point loop Sling...

Страница 8: ...centre line outside facing the outside The sling label is located on the outside WARNING To avoid injury during log roll make sure there is another caregiver present and or use the side rails on the o...

Страница 9: ...Make sure that all loops are attached securely and the patient lays comfortably in the sling Fig 6 Fig 7 Fig 8 Fig 9 Turning to the Side Follow these 9 steps 1 Position the spreader bar in line with...

Страница 10: ...lays comfortably in the sling 6 If an adjustment is needed lower the patient and make sure that the weight of the patient is taken up by the receiving surface before removing the loops 7 If possible...

Страница 11: ...ig 15 5 Lower the spreader bar to enable the straps to be attached Caregiver should always hold on to the spreader bar to avoid swinging movements See Fig 16 6 Attach the loops Use the loop lengths th...

Страница 12: ...e Repositioning Sling make sure the Repositioning Sling is not attached to any other object than the spreader bar make sure that there is clearance to perform the manoeuvre 11 If an adjustment shiftin...

Страница 13: ...g to local hygiene legislation If allowed according to the product label tumble dry in low temperature max 60 C 140 F 4 Do NOT wash with other items that have rough surfaces or sharp objects use any m...

Страница 14: ...tress or pressure or to excessive heat or humidity The slings should be kept away from sharp edges corrosives or other things that could cause damage on the sling Service and Maintenance It is recomme...

Страница 15: ...See Sling Selection on page 6 Operating Transport and Storage Environment Temperature 0 C to 40 C 32 F to 104 F Operating and Storage 25 C to 60 C 13 F to 140 F Transport Humidity 15 70 at 20 C 68 F...

Страница 16: ...ulation 2017 745 AHD001 Symbols Do NOT wash symbol Located on the outside of the sling Do NOT use lift symbol The AHD001 has been washed Located on the outside of the sling Symbol for Loop Use a loop...

Страница 17: ...Intentionally left blank...

Страница 18: ...Intentionally left blank...

Страница 19: ...oor HQ Building North Tower Dubai Science Park Al Barsha South P O Box 11488 Dubai United Arab Emirates Direct 971 487 48053 Fax 971 487 48072 Email Info ME arjo com NEDERLAND Arjo Nederland BV Biezen...

Страница 20: ...ensure ergonomic patient handling personal hygiene disinfection diagnostics and the effective prevention of pressure ulcers and venous thromboembolism we help professionals across care environments to...

Отзывы: