59
Mains disconnection device
Permanently installed in wall. Visible and accessible at all times
Mains transient voltage
4 kV*
Creepage distance
3 mm*
Air clearance
1.8 mm*
Direction of movement
Up and down, right and left,
clockwise and counterclockwise
* If local regulations are higher, then they have to be met.
Water supply
Cold/Warm Water
3/4” B.S.P female
Connect with shut off valves between unit and supply net, minimum through Ø 16 mm
Maximum static pressure
6 bar (600 kPa, 87 psi)
Minimum operating pressure (all markets except UK market)
Minimum operating pressure (UK market)
2 bar (200 kPa, 29 psi)
1 bar (100 kPa, 14.5 psi)
Required cold water temperature
2 - 20 °C (36 - 68 °F)
Required warm water temperature
60 - 80 °C (140 - 176 °F)
Drain position
Floor or wall
Floor
Maximum floor load
990 kg /0.91 m
2
Maximum floor pressure
1.8 N/mm
2
Operating conditions
Ambient temperature
40 °C (104 °F)
Relative humidity
30% - 75% Rh
Atmospheric pressure
800 hPa to 1060 hPa
Transport and storage
Ambient temperature
-20 °C to 70 °C
(-4 °F to 158 °F)
Relative humidity
10% - 80% at 20 °C (68 °F)
Atmospheric pressure
500 hPa to 1100 hPa