J405 TROUBLESHOOTING
Symptom
Possible Cause
Solution
Power Problems
Instrument does not turn ON.
Battery will not hold a charge.
Discharged battery or
Dead battery.
Recharge battery for a minimum of
Replace battery. Refer to page
Instrument does not turn on when
connected to AC power
supply/charger.
Open fuse.
Insufficient power.
Internal component failure.
Wait for fuse to automatically reset
itself.
Use a voltmeter to verify there is
power to the AC outlet.
Call AZI Customer Service for
information at 800-528-7411 or
602-470-1414.
Sampling Problems
Airflow is restricted during the
sensor regeneration cycle, causing
possible permanent damage.
Kinks and crimps in the
Tygon
®
tubing.
Obstructed intake or exhaust.
Clogged intake filter
Periodically check the Tygon
®
tubing in the instrument. Refer to
page
Verify ports are not obstructed.
Change intake fritware filter. Refer
High erratic results.
Internal mercury vapor
contamination.
1.
Install Zero Air Filter in intake.
Press SAMPLE button. After
three samples, if readings are
over 2 μg/m
3
, replace fritware
filter and Tygon
®
tubing. Refer
2.
Perform sensor regeneration
with the Zero Air Filter in
intake. Refer to page 23. Retest
if necessary. Replace scrubber
filters and Tygon
®
tubing. Refer
High or erratic results
Intake and internal filters may
become clogged and need
replacement when sampling in a
dusty area.
1.
Open instrument and check for
pinched, crimped or
disconnected internal tubing.
2.
In extreme conditions, an
additional particle filter may be
installed on the intake.