241434C
Ranger Blood/Fluid Warming
English 7
Warming units Rev A to M
Independent back-up
safety alarm - 46°C
Alarm indicator light
•
illuminates and alarm
sounds, alphanumeric
display alternately
flashes a temperature of
43°C or higher and the
word “HI�”
Alphanumeric display
•
is dark, alarm sounds
(backup safety alarm
still works even if
display is dark)�
Heater
temperature
rose to 46°C�
Safety backup
system is
activated at
46°C and unit
shuts off power
to the heating
plates�
TURN OFF THE
UNIT AND UNPLUG
IT� Do not use the unit�
Discard the disposable
set� Call Arizant
Healthcare Customer
Service�
Warming units Rev N and newer
Independent back-up
safety alarm - 44°C
Alarm indicator light
•
illuminates and alarm
sounds, alphanumeric
display alternately
flashes a temperature of
43°C or higher and the
word “HI�”
Alphanumeric display
•
is dark, alarm sounds
(backup safety alarm
still works even if
display is dark)�
Heater
temperature
rose to 44°C�
Safety backup
system is
activated at
44°C and unit
shuts off power
to the heating
plates�
TURN OFF THE
UNIT AND UNPLUG
IT� Do not use the unit�
Discard the disposable
set� Call Arizant
Healthcare Customer
Service�
*Transient conditions may cause an over-temperature alarm condition� These conditions include:
There was an extreme change in flow rates (e�g�, from 500 mL/min to stop flow)�
•
The unit was turned on and reached the set point temperature of 41°C before the warming
•
cassette was inserted in the device�
Fluids were prewarmed to above 42°C before being infused�
•