14 English
Ranger Blood/Fluid Warming
Arizant Healthcare Inc.
Specifications
Physical characteristics
warming
unit
4�5 in� (11 cm) high x 7�5 in� (19 cm) wide x 10 in� (25 cm) long; wt�: 7 lb� 7 oz� (3�4 kg)
classification
Classified under IEC 601-1 Guidelines (and other national versions of the Guidelines) as Class I,
Type B, Ordinary Equipment� Not suitable for use in the presence of flammable anesthetic mixtures
with air or with oxygen or nitrous oxide� Classified by Underwriters Laboratories Inc� with respect to
electric shock, fire, and mechanical hazards only, in accordance with UL 2601-1 and in accordance
with Canadian/CSA C22�2 No� 601�1�
Leakage current
Meets leakage current requirements in accordance with UL 60601-1 and IEC 601-1�
Environmental conditions
operating
temperature
range
15° to 40°C (59° to 104°F)
storage
temperature
range
-20° to 45°C (-4° to 113°F)
operating
humidity
10 to 85% RH, noncondensing
atmospheric
pressure
range
50 kPa to 106 kPa
Electrical characteristics
Temperature characteristics
device
rating
100-120 VAC, 50/60 Hz
220-240 VAC, 50/60 Hz
set
point
temperature
41°C
maximum
heating
power
900 W
over
-
temperature
alarm
43°C
fuse
10 Amp
under
-
temperature
alarm
33°C
over
-
temperature
cutoff
44°C (warming units Rev N and newer)
46°C (warming units Rev A to M)