Ranger Irrigation Fluid Warming System
English
Model 247 Product Safety Features
The following chart describes the safety alarm features of the Ranger Irrigation warming unit.
note
:
The difference in over-temperature alarm system cut-off points does not indicate any difference
in safety and is simply a result of tighter controller tolerances.
The Ranger Irrigation Fluid Warming System is designed with very tight controls to regulate its temperature
to the 41°C set-point. The microprocessor controller measures the temperature at the fluid pathway inside the
warming plate four times every second, making it very responsive to changes in flow rate.
ALARM TyPE
SET-POINT
INDICATION
DESCRIPTION
Primary over-temperature
alarm system
48°C
LED alternately displays
“HI” and temperature,
alarm sounds, alarm light
flashes.
Primary Alarm: Alarm
sounds, controller cuts
power to the heaters until
temperature falls below 48°C.
When heater temperature
drops below 48°C, audible
and visual alarms stop.
Independent back-up
over-temperature alarm
system (cut-off).
50°C
LED displays the
temperature or has gone
dark, alarm sounds.
Primary failure. Temperature
will rise to 50°C within a few
seconds. The safety backup
system is activated at 50°C
and power is immediately shut
off to the
heating plates.
Under-temperature alarm
33°C
LED displays “LO”
and temperature, alarm
sounds, alarm light
flashes.
Heater temperature has fallen
to 33°C. Alarms will stop
when temperature rises above
33°C. Continue use of unit.
ALARM TyPE
SET-POINT
INDICATION
DESCRIPTION
Primary over-temperature
alarm system
48°C
LED alternately displays
“HI” and temperature,
alarm sounds, alarm light
flashes.
Primary Alarm: Alarm
sounds, controller cuts
power to the heaters until
temperature falls below 48°C.
When heater temperature
drops below 48°C, audible
and visual alarms stop.
Independent back-up
over-temperature alarm
system (cut-off).
50°C
LED displays the
temperature or has gone
dark, alarm sounds.
Primary failure. Temperature
will rise to 50°C within a few
seconds. The safety backup
system is activated at 50°C
and power is immediately shut
off to the
heating plates.
Under-temperature alarm
33°C
LED displays “LO”
and temperature, alarm
sounds, alarm light
flashes.
Heater temperature has fallen
to 33°C. Alarms will stop
when temperature rises above
33°C. Continue use of unit.
Содержание Bair Hugger Ranger 247
Страница 14: ...Deutsch 13 ...
Страница 26: ...Français 27 ...
Страница 38: ...Italiano 41 ...
Страница 50: ...Español 55 ...
Страница 62: ...Português 69 ...
Страница 74: ...Nederlands 83 ...
Страница 86: ...Dansk 97 ...
Страница 98: ...Svenska 111 ...
Страница 110: ...Norsk 125 ...
Страница 122: ...Suomi 139 ...
Страница 134: ...Ελληνικά 153 ...
Страница 146: ...日本語 167 ...
Страница 157: ...レンジャーイリゲーション液加温システムは 日本語 180 記号の定義 オン オフ ヒューズ 注意 添付文書類を参照してください タイプ B 適用部品 電圧 交番電流 AC ...
Страница 158: ...Polski 181 ...
Страница 170: ...Magyar 195 ...
Страница 182: ...Česky 209 ...