Subject to change · V 4.06.19 · EN
Page 44
www.argo-hytos.com
17. Technical data
1
Depending on the oil viscosity
2
Depending on the permissible operating pressure
3
From software version 1.70.15 upwards
4
Calibrated to n-Pentan at 25 °C (77 °F)
5
Calibrated to air at room temperature
Parameter
Size
Unit
Operating pressure
High-pressure connection
1
With pump operation
10 ... 320 (145 ... 4,600)
0
bar (psi)
bar (psi)
Viscosity range fluid
2
5 ... 1000
mm²/s
Operating temperature range fluid
0 ... +60 (+32 ... +140)
°C (°F)
Operating conditions
Temperature
Rel. humidity
-10 ... +60 (+14 ... +140)
0 ... 95
°C (°F)
% r.H. (non-condensing)
Compatible fluids
mineral oils (H, HL, HLP, HLPD, HVLP),
synthetic esters (HETG, HEPG, HEES, HEPR),
polyalkylenglycols (PAG),
zinc and ash-free oils (ZAF),
polyalphaolefins (PAO)
Wetted materials
chrome, aluminum, stainless steel, Viton, steel, brass, HNBR, NBR,
polyurethane resin, epoxy resin, chemical nickel/gold (ENIG),
soldering tin (Sn96, 5Ag3CuO, 5NiGe), aluminum oxide, glass
(DuPont QQ550), gold, silver-palladium, sapphire, PVC (hoses)
Power supply device
Power supply
Power consumption
24
max. 8
VDC
A
Power supply for the according power adaptor
Power supply
Power consumption
Power at 24VDC-output
100 ... 240
max. 4
max. 221
VAC (50/60 Hz)
A
W
Characteristics battery
Nominal capacity
Loading time
Running time when measuring without pump
(when measuring with pump the running time
decreases depending on the oil viscosity)
7500
< 1
> 24
mAh
h
h
Display particle measurement
ISO 4406:99
SAE AS 4059E
NAS 1638 (based)
3
GOST 17216 (based)
3
Size channels
0 ... 28 (calibrated area 10 ... 22)
000 ... 12
00 ...12
00 ...17
4, 6, 14, 21
ordinal number (OZ)
ordinal number (OZ)
ordinal number (OZ)
ordinal number (OZ)
μm(c)
Measuring range oil parameter
Rel. permittivity
Rel. humidity
Conductivity
Temperature
1 ... 7
0 ... 100
100 ... 800000
-20 ... +120 (-4 ... +248)
-
%
pS/m
°C (°F)
Measuring accuracy
Particle measurement (within calibr. range) - ISO 4 / ISO 6
Particle measurement (within calibr. range) - ISO 14 / ISO 21
Rel. dielectric number
4
Rel. humidity (10 ... 90%)
5
Rel. humidity (<10 %, >90%)
5
Conductivity (100 ... 2000 pS/m)
Conductivity (2000 ... 800000 pS/m)
Temperature
± 1
± 2
± 0.015
± 3
± 5
± 200
Typ. < 10
± 2
ordinal number (OZ)
ordinal number (OZ)
-
% r.H.
% r.H.
pS/m
%
K