ARGUS service
14.227._A60_A100 M en.SM.V4.3X
ARGUS 28.03.06 / PJ
7 / 22
Select the next step by pressing one of the
buttons (“Configuration” or “View history”).
4.2. ARGUS service – Configuration
Important remark:
After each configuration change,
a functional check has to be done!
4.2.1. Docking configuration
All configuration possibilities (indexes) mentioned in
chapter 2.2
can be modified
herein in its own tree structure as shown above.
All indexes which are different from the ARGUS firmware default are highlighted.
CAUTION!
Every pump used with the ARGUS 60 M or ARGUS 100 M docking station has to be
preconfigured with its corresponding serial number label!
Index #396 in the configuration of each pump must be programmed with the corre-
sponding serial number. Please refer to the pumps respective service manual to
change these indexes if necessary. In the case of an invalid serial number, the dock-
ing station will generate an alarm and the pump can not be programmed via barcode.