ARGUS service
14.227._A60_A100 M en.SM.V4.3X
ARGUS 28.03.06 / PJ
6 / 22
4. ARGUS
SERVICE
4.1. General
The new ARGUS
service
utility is a highly user friendly PC software which can config-
ure and upgrade ARGUS devices over PC serial COM port. With this Windows based
software you can also set pump clock, change PIN code, read and print out history
and easily replicate pump configurations, and so on. The modern and clearly struc-
tured design of this self-describing PC-tool allows a very easy and rapid modification
of the ARGUS 600 Syringe pump, the ARGUS 707 / 708 Volumetric infusion pump
and the ARGUS 60 M / 100 M docking station. This software may be available from
your local distributor or directly from ARGUS Medical.
REMARK:
“ARGUS
service
” may only be used with V4 software versions greater or equal to 4.30.
CAUTION!
The docking station has to be disconnected from the pumps before and while the se-
rial interface cable is connected to the docking station.
Connect a straight RS-232 cable from the docking data interface to the serial in-
terface (COM port) of your PC.
Start the ARGUS
service
Press the ‘
Docking Station
’ button.
Select “Start Configurator” (
see next chapter
)
or “Start Update-Center” (
see chapter 5.5.
).