Burn Navigator® User’s Manual
Page 43
Cause of Alert or Message
Alert or Message Text
The user presses the Hourly Update
button, but it is too late to redo the last
hourly update and it is not yet time for
a new hourly update.
“It is not yet time to do the hourly fluid update or
the 10 minute edit window has expired.
To update the pump setting, go to the Home screen
and select the “Current Infusion Rate” field.”
During a wireless handoff, the system
time on the receiver is more than 30
minutes ahead or behind of the device
that exported the patient file
Warning – Time Offset
A time difference of __ minutes has been detected.
Please check that your device’s date and time are
set correctly before proceeding.
If you continue, the received patient data may be
adjusted to correct for the difference.
Go ahead with this transfer?
When user selects “yes” for presence of
gross myoglobinuria. [This message
shows on the recommendation screen]
ABLS guidelines recommend starting at
4mL/kg/TBSA for adults with electrical injuries.
On the new patient setup,
recommendation screen: When the
initial formula exceeds 2,000 mL/hr.
Initial formula exceeds 2,000 mL/hr.
Recommending 2,000 mL/hr.
On the new patient setup,
recommendation screen: When the
initial formula is below 100 mL/hr.
Initial formula is below 100 mL/hr.
Recommending 100 mL/hr.
On the fluid update wizard: When
increasing the recommended infusion
rate would cause the patient’s projected
24 hour volume to be too high.
High 24 hour volume projected! Fluid increase not
recommended.
On the fluid update wizard: When
maintaining the current infusion rate
would cause the patient’s projected 24
hour volume to be too high.
High 24 hour volume projected! Fluid decrease
recommended.