Specifications
49
4.2 Calculated Parameters
Volt-Amps
Calculated as the product of the rms voltage and the rms current.
Range:
0.004 VA to 20 kVA (with supplied 1000:1 current
transformer)
Accuracy:
Derived from the individual uncertainties of the
voltage and current measurements.
±
1.5% of reading, max.
Volt-Amps Reactance
Calculated as the product of the rms voltage, the rms current, and
the sine of the phase angle between the two.
Range:
0.004 var to 20 kvar (with supplied 1000:1 current
transformer)
Accuracy:
Derived from the individual uncertainties of the
voltage, current, and phase measurements
±
1.5% of VA, max.
Power
Calculated as the product of the rms voltage, the rms current, and
the cosine of the phase angle between the two.
Range:
0.004 W to 20 kW (with supplied 1000:1 current
transformer)
Accuracy:
Derived from the individual uncertainties of the
voltage, current, and phase measurements
±
1.5% of VA, max.