PL7004 Kit
32
Rev F
4.4.2
Cables
Periodically inspect fiber optic cables for physical damage. Replace damaged cables.
Do not manually force
connector shutters open. Damage to the connectors will result and laser radiation may escape.
4.4.3
Probe Interface
Periodically inspect the interface for physical damage. There are no user-repairable parts in the interface.
The laser module in the interface is factory-replaceable with one of the same type without requiring re-
calibration of the probe kit. The laser used in the interface will gradually degrade in light (optical) output over
its operating life. Noticeable degradation is likely in about 10,000 operating hours. The unit as delivered has
adequate optical power margin to allow for some degradation under specified operating conditions. The light
output reduction prior to noticeable performance degradation varies by model and application conditions and
is affected by the number and length of fiber cable extensions.
If damage to the interface is observed, return to the factory for repair.
4.5 RETURN PROCEDURE
To return a probe to AR, use the following procedure:
1.
Contact the AR Customer Service Department.
2.
Briefly describe the problem in writing. Give details regarding the observed symptom(s), and whether the
problem is constant or intermittent in nature. If AR has been previously contacted about the problem,
provide the date(s), identify the contact and the nature of the contact. Include the serial number of the
item being returned along with the Return Authorization (RMA) number obtained from the AR Customer
Service Department.
NOTE: Returns from outside the United States are not permitted without prior
authorization. If shipping from outside of the United States, closely follow all directions on the RMA
form for return shipping and marking.
3.
Package the probe carefully. Use the original boxes and packing materials, if possible.
4.
Mail the package to:
AR RF/Microwave Instrumentation
160 Schoolhouse Rd
Souderton PA USA 18964
Attn: Customer Service
If the probe is still under warranty, refer to the Limited Warranty for additional information about your
return.
4.6 PERIODIC CALIBRATION
The frequency of calibration and level (grade) of calibration should be determined by the user after carefully
considering the application requirements. Normal considerations include field accuracy requirements, test
environment, mode of use, frequency range and field level range. Users may wish to refer to IEEE Std 1309-
2003 for guidelines in specifying and selecting the calibration details. This standard includes consensus
calibration methods for many different field probes, including recommendations for probe types that are
similar to this probe kit.
Prior to returning a probe kit for calibration, the user should specify what type of calibration is requested,
standard or custom, and request a quote for the calibration. User can specify normal AR calibration (standard)
Содержание FI7000
Страница 2: ......
Страница 4: ......
Страница 5: ......
Страница 6: ......
Страница 20: ......
Страница 24: ......
Страница 28: ...PL7004 Kit 12 Rev F...
Страница 46: ...PL7004 Kit 30 Rev F...
Страница 50: ...PL7004 Kit 34 Rev F...
Страница 51: ...Rev F 35 Appendix A Dimensional Drawings Figure A 1 Model PL7004...
Страница 52: ...PL7004 Kit 36 Rev F...
Страница 54: ......